FDA Adverse Event Malfunction Summary report: N

KNIFE MCO5D MICROFRANCE SUCTION RND 2MM

MDR report key: 8252842 · Received January 16, 2019

Report

Report Number
2523190-2019-00003
Event Type
Malfunction
Date Received
January 16, 2019
Date of Event
December 6, 2018
Report Date
December 26, 2018
Manufacturer
INTEGRA MICROFRANCE S.A.S.
Product Code
JYO
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE DEVICE HISTORY RECORD (DHR ) SHOWS NO ANOMALIES THAT COULD BE ASSOCIATED WITH THE COMPLAINT. THEREFORE, AN INVESTIGATION CAN NOT BE CARRIED OUT BECAUSE THE INSTRUMENT OF THE COMPLAINT HAS NOT BEEN RETURNED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D10, G4, G7, H2, H3, H6, H10. THE DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS BROKEN AT THE WELDING AREA. THE TUBE WITH THE ACTIVE PART WAS SLIGHTLY BENT. THE CLEANING WIRE WAS MISSING. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES THAT COULD BE ASSOCIATED WITH THE COMPLAINT WERE OBSERVED. THE REPORTED COMPLAINT WAS CONFIRMED. THE BREAKAGE WAS PROPABLY DUE TO A BAD HANDLING OF THE DEVICE.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT YET RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE PLANT INVESTIGATION IS IN PROGRESS AND A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A DOCTOR REPORTED THAT ON (B)(6) 2018 WHILE ADJUSTING THE SHAFT OF THE MCO5D KNIFE MICROFRANCE SUCTION RND, DURING FIRST TIME USE, THERE WAS A FRACTURE IN THE CONNECTION BETWEEN TWO PARTS. A (B)(6) MALE PATIENT WAS PREPPED FOR AN UNSPECIFIED SURGERY. THERE WAS NO PATIENT CONTACT, PATIENT INJURY/DEATH ALLEGED. THERE WAS APPROXIMATELY 0.5 HOUR (30 MINUTES) OF DELAY IN SURGERY. ADDITIONAL INFORMATION RECEIVED ON 28DEC2018 STATED THAT THE SURGICAL DELAY DID NOT CAUSE AN IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
44846 KNIFE MCO5D MICROFRANCE SUCTION RND 2MM PFM11 JYO INTEGRA MICROFRANCE S.A.S. 2789466

Patients

Seq Age Sex Outcome Treatment
1 46 YR