KNIFE MCO5D MICROFRANCE SUCTION RND 2MM
Report
- Report Number
- 2523190-2019-00003
- Event Type
- Malfunction
- Date Received
- January 16, 2019
- Date of Event
- December 6, 2018
- Report Date
- December 26, 2018
- Manufacturer
- INTEGRA MICROFRANCE S.A.S.
- Product Code
- JYO
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE DEVICE HISTORY RECORD (DHR ) SHOWS NO ANOMALIES THAT COULD BE ASSOCIATED WITH THE COMPLAINT. THEREFORE, AN INVESTIGATION CAN NOT BE CARRIED OUT BECAUSE THE INSTRUMENT OF THE COMPLAINT HAS NOT BEEN RETURNED.
ADDITIONAL INFORMATION: D10, G4, G7, H2, H3, H6, H10. THE DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS BROKEN AT THE WELDING AREA. THE TUBE WITH THE ACTIVE PART WAS SLIGHTLY BENT. THE CLEANING WIRE WAS MISSING. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES THAT COULD BE ASSOCIATED WITH THE COMPLAINT WERE OBSERVED. THE REPORTED COMPLAINT WAS CONFIRMED. THE BREAKAGE WAS PROPABLY DUE TO A BAD HANDLING OF THE DEVICE.
THE DEVICE WAS NOT YET RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE PLANT INVESTIGATION IS IN PROGRESS AND A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
A DOCTOR REPORTED THAT ON (B)(6) 2018 WHILE ADJUSTING THE SHAFT OF THE MCO5D KNIFE MICROFRANCE SUCTION RND, DURING FIRST TIME USE, THERE WAS A FRACTURE IN THE CONNECTION BETWEEN TWO PARTS. A (B)(6) MALE PATIENT WAS PREPPED FOR AN UNSPECIFIED SURGERY. THERE WAS NO PATIENT CONTACT, PATIENT INJURY/DEATH ALLEGED. THERE WAS APPROXIMATELY 0.5 HOUR (30 MINUTES) OF DELAY IN SURGERY. ADDITIONAL INFORMATION RECEIVED ON 28DEC2018 STATED THAT THE SURGICAL DELAY DID NOT CAUSE AN IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 44846 | KNIFE MCO5D MICROFRANCE SUCTION RND 2MM | PFM11 | JYO | INTEGRA MICROFRANCE S.A.S. | 2789466 |