FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 8252458 · Received January 16, 2019

Report

Report Number
1000113657-2019-00036
Event Type
Malfunction
Date Received
January 16, 2019
Date of Event
December 20, 2018
Report Date
January 16, 2019
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00041220001286
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20-USER'S TEST STRIP HAD POOR STORAGE(KITCHEN). TEST STRIP UDI (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER (SEVERAL ATTEMPTS) IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH THE CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CUSTOMER TO CONTACT CUSTOMER CARE.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 233, 155 AND 383 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80 - 106 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON 12/20/2018, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 04/30/2019 AND OPEN VIAL DATE IS (B)(6) 2018. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1:76MG/DL DATE: (B)(6) 2018 TIME:6:26AM FASTING RESULT 2:233MG/DL DATE: (B)(6) 2018 TIME:6:15AM FASTING. RESULT 3:155MG/DL DATE: (B)(6) 2018 TIME:4:52PM FASTING. RESULT 4:383MG/DL DATE: (B)(6) 2018 TIME:8:46PM FASTING. RESULT 5:501MG/DL DATE: (B)(6) 2018 TIME:8:45PM NON-FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
45704 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MU2666 00041220001286

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY