FDA Adverse Event
Malfunction
Summary report: N
ANIMAS VIBE
MDR report key: 8252198
·
Received January 16, 2019
Report
- Report Number
- 2531779-2019-00448
- Event Type
- Malfunction
- Date Received
- January 16, 2019
- Report Date
- January 9, 2019
- Manufacturer
- ANIMAS LLC
- Product Code
- OYC
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/09/2019 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE AUDIO BOLUS BUTTON COVER WAS LIFTED. THE AUDIO BOLUS BUTTON WAS INOPERABLE DUE TO THE AUDIO BOLUS BUTTON SLUG MISSING. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE AUDIO BOLUS BUTTON WAS INOPERABLE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 01/09/2019.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47285 | ANIMAS VIBE | INSULIN INFUSION PUMP | OYC | ANIMAS LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |