FDA Adverse Event
Malfunction
Summary report: N
TANDEM T:SLIM INSULIN DELIVERY SYSTEM
MDR report key: 8250953
·
Received January 15, 2019
Report
- Report Number
- 3013756811-2019-02073
- Event Type
- Malfunction
- Date Received
- January 15, 2019
- Date of Event
- October 1, 2018
- Report Date
- January 15, 2019
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- LZG
- PMA / PMN Number
- K111210
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INVESTIGATION HAS BEEN COMPLETED. BASED ON THE ANALYSIS, AN ADDITIONAL RESULT WAS IDENTIFIED (170) AND A DIFFERENT ISSUE WAS IDENTIFIED (4116).
Additional Manufacturer Narrative · 1
PUMP SERIAL NUMBERS: (B)(4). CARTRIDGE LOT NUMBER: M021850. USER OVERFILLING THE CARTRIDGE BAG CAUSING A PRESSURE LEAK.
Description of Event or Problem · 1
QUARTERLY SUMMARY REPORT FOR ALLEGED ALTITUDE ALARMS: IT WAS REPORTED THAT AN ALTITUDE ALARM OCCURRED WITHIN LABELED ALTITUDE RANGE. THE ALARM WAS ULTIMATELY CLEARED OR USER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY TO ADDRESS THE ISSUE. THERE WAS NO ADVERSE EFFECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 41214 | TANDEM T:SLIM INSULIN DELIVERY SYSTEM | INSULIN PUMP | LZG | TANDEM DIABETES CARE | 004628 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |