BBL SENSI-DISC TICARCILLIN TIC-75
Report
- Report Number
- 1119779-2007-00003
- Event Type
- Malfunction
- Date Received
- February 23, 2007
- Date of Event
- January 30, 2007
- Report Date
- February 19, 2007
- Manufacturer
- BD CARIBE, LTD.
- Product Code
- JTN
- Removal / Correction Number
- 1119779-02/19/2007-R001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INTERNAL INVESTIGATION OF RETENTION PRODUCT ALSO EXHIBIT A FALSE SUSCEPTIBLE RESULT WHEN TESTED AGAINST E. COLI ATCC 35218. INTERNAL INVESTIGATION HAS SHOWN THAT THIS LOT OF TICARCILLIN CONTAINS CLAVULANIC ACID, A COMPOUND THAT INACTIVATES E. COLI ATCC 35218'S BETA-LACTAMASE ENZYMES, ALLOWING THE ANTIBIOTIC TO KILL THE ORGANISM CREATING A ZONE OF GROWTH INHIBITION WITH THE KIRBY-BAUER TEST METHODOLOGY.
CUSTOMER WAS HAVING PROBLEMS WITH THE DISCS. CUSTOMER HAS BEEN PERFORMING QC WITH E. COLI 25922 AND P. AERUGINOSA AND HAD ACCEPTABLE RESULTS. CUSTOMER STARTED NOTICING SOME PATIENT ISOLATES, MAINLY K. PNEUMO., THAT WERE SENSITIVE TO TICARCILLIN, WHEN CUSTOMER KNEW THEY SHOULD BE RESISTANT. CUSTOMER TESTED ISOLATES ON VITEK, AND VITEK CALLED THEM RESISTANT. CUSTOMER THEN SET UP E. COLI 35218, AND GOT A ZONE SIZE OF 20, WHEN IT SHOULD BE 6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BBL SENSI-DISC TICARCILLIN TIC-75 | ANTIMICROBIAL SUSCEPTIBILITY TEST DISC | JTN | BD CARIBE, LTD. | * | 6167435 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Life Threatening |