FDA Adverse Event Malfunction Summary report: N

3.0 RIO® ROBOTIC ARM - MICS

MDR report key: 8248299 · Received January 15, 2019

Report

Report Number
3005985723-2019-00058
Event Type
Malfunction
Date Received
January 15, 2019
Date of Event
January 3, 2019
Report Date
March 26, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486030407
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

REPORTED EVENT: AN EVENT REGARDING INCORRECT GREEN VISUAL REMOVAL DURING A TOTAL KNEE PROCEDURE INVOLVING 3.0 RIO ROBOTIC ARM - MICS, CATALOG: 209999 WAS REPORTED. METHOD & RESULTS: DEVICE HISTORY REVIEW: A REVIEW OF THE DHR ASSOCIATED WITH RIO 29 FOUND QUALITY INSPECTION PROCEDURES SUCCESSFULLY PASSED. COMPLAINT HISTORY: BASED ON THE DEVICE IDENTIFICATION (PN 209999) THE COMPLAINT DATABASES WERE REVIEWED FROM 2011 TO PRESENT FOR SIMILAR REPORTED EVENTS REGARDING GREEN AREA NOT REMOVING DURING CUTTING. THERE WERE OTHER REPORTED EVENTS FOR THE LISTED CATALOG NUMBER (B)(4). CONCLUSION: RESOLUTION PER (B)(4): SUBCASE (B)(4) NOTES 1/6/2019 21:08:15 MVINSH ACTION TYPE: DESCRIPTION MV 1/4/2019. (B)(4). MPS (B)(6) REPORTED GREEN RESECTION AREA NOT DISAPPEARING DURING CUTTING. SUBCASE (B)(4) NOTES 1/6/2019 21:16:49 MVINSH ACTION TYPE: RESOLUTION MV 1/4/2019. (B)(6). (B)(4). PKA 2.5.6.4. THA 3.1.1.1. TKA 1.0. CRISIS 2.13.1.2K/2.13.1.1/2.13.2. MGO 1.1.8. UTILITIES 1.10. (B)(4) GREEN RESECTION AREA NOT DISAPPEARING DURING CUTTING. REPLACED CPCI ASSEMBLY. NOTICED CAMERA WAS LOOSE, TENSIONED SUPPORT SCREW. -CALIBRATION END EFFECTOR WAS NOT SEATED FIRM ONTO J6 AFTER TIGHTENING. REPLACED J6 SCREW INSERTS AND CAL EE SCREWS. ALL TESTING COMPLETE. SYSTEM IS READY FOR CLINICAL USE.

Description of Event or Problem · 0

(B)(4) - MPS (B)(6) REPORTED GREEN RESECTION AREA NOT DISAPPEARING DURING CUTTING DESCRIPTION OF EVENT: NO GREEN DISAPPEARING WHILE CUTTING. APPLICATION: TKA. CASE DELAY > 15.

Additional Manufacturer Narrative · 1

¿AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.¿

Description of Event or Problem · 1

(B)(4) - (B)(6) REPORTED GREEN RESECTION AREA NOT DISAPPEARING DURING CUTTING. DESCRIPTION OF EVENT: NO GREEN DISAPPEARING WHILE CUTTING. APPLICATION: TKA. CASE DELAY: >15.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
43242 3.0 RIO® ROBOTIC ARM - MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 00848486030407

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization