MENTOR SMOOTH ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2019-07937
- Event Type
- Injury
- Date Received
- January 15, 2019
- Date of Event
- November 24, 2018
- Report Date
- December 21, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ON 2/25/2019, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IDENTIFIED AS CATALOG NUMBER 3501690, AND LOT NUMBER 5904883. A MANUFACTURING RECORD EVALUATION (MRE) REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. SINCE THE DEVICE RECEIVED, ON 2/25/2019, MENTOR ESTIMATED THE DATE OF EXPLANTATION ON (B)(6)2019, UNTIL FURTHER INFORMATION. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
DEVICE EVALUATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE CONTAINED NO FLUID. WHITE MATERIAL WAS OBSERVED WITHIN THE DEVICE AND NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. DURING INITIAL EVALUATION SOME CREASE WERE OBSERVED ON THE ANTERIOR AND POSTERIOR ASPECT. LEAK TESTING WAS PERFORMED ACCORDING WITH MENTOR PROCEDURES AND IT REVEALED A RENT WITHIN SHELL ABRASION ON THE POSTERIOR ASPECT, MEASURING APPROXIMATELY <0.1 CM. NO OTHER ANOMALIES WERE OBSERVED. DEFLATION COMPLAINT WAS CONFIRMED. MENTOR PRODUCTS ARE 100% VISUALLY INSPECTED PRIOR TO RELEASE IN ADDITION TO THOROUGH IN-PROCESS TESTING DURING SEVERAL STAGES OF THE MANUFACTURING PROCESS. MENTOR PRODUCT ANALYSIS LAB CONCLUDED THAT THE RENT OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. THE COMPLAINT WAS CONFIRMED SINCE A RENT WITHIN A SHELL ABRASION WAS FOUND ON THE DEVICE. THESE FINDINGS ARE CONSISTENT WITH A CREASE/FOLD FAILURE WHICH IS A KNOWN INHERENT RISK ASSOCIATED WITH THE USE OF SALINE-FILLED MAMMARY PROSTHESES AND MAY BE THE RESULT OF ONE OR MORE OF THE FOLLOWING CONTRIBUTING FACTORS: UNDERINFLATION OR OVER INFLATION OF THE DEVICE, CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE - SUCH AS CAPSULAR CONTRACTURE OR TOO SMALL A BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. ALL THE IMPLANTS ARE 100% INSPECTED BEFORE LEAVING THE FACILITY. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON (B)(6) 2019, THE MANUFACTURING RECORD EVALUATION (MRE) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO DEVICE HISTORY RECORD (DHR) REVIEW COULD BE PERFORMED. CONCOMITANT PRODUCTS: RIGHT-UNKNOWN SALINE IMPLANT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT (B)(6) YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH UNKNOWN SALINE BREAST IMPLANTS WHICH THE LEFT SIDE DEFLATED AFTER IMPLANTATION. PATIENT NOTICED SHE HAD A DEFLATION ON THE LEFT SIDE BECAUSE IT WAS NOT AS FULL AS THE RIGHT SIDE. DOCTOR CONFIRMED LEFT DEFLATION ON (B)(6) 2018. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 42150 | MENTOR SMOOTH ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 5904883 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Other| R |