ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE
Report
- Report Number
- 3001845648-2019-00018
- Event Type
- Malfunction
- Date Received
- January 15, 2019
- Date of Event
- November 8, 2018
- Report Date
- December 18, 2018
- Manufacturer
- COOK IRELAND LTD
- Product Code
- FCG
- UDI-DI
- 00827002315204
- PMA / PMN Number
- K083330
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
510 (K) NUMBER; K083330. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G. 1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(4) COOK MEDICAL INCORPORATED (CMI) 1025 ACUFF ROAD P.O BOX 4195 BLOOMINGTON INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: 3005580113. DEVICE EVALUATION COMPLAINT DEVICE WAS NOT RETURNED THEREFORE A DOCUMENT BASED REVIEW WAS PERFORMED. DOCUMENT REVIEW PRIOR TO DISTRIBUTION, ALL ECHO-19 DEVICES ARE SUBJECTED TO FUNCTIONAL CHECKS AND VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR ECHO-19 OF LOT NUMBER C1457029 DID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. THE REVIEW OF RELEVANT MANUFACTURING RECORDS, CONFIRMS THE FAILURE MODE HAS NOT PREVIOUSLY OCCURRED WITH THE CURRENT LOT NUMBER. BASED ON THE INFORMATION AVAILABLE TO DATE, THERE IS NO EVIDENCE TO SUGGEST THAT THERE ARE ANY MANUFACTURING ISSUES ASSOCIATED WITH LOT NUMBER C1457029. IFU REVIEW THE NOTES SECTION OF THE INSTRUCTIONS FOR USE, IFU0101-0 WHICH ACCOMPANIES THIS DEVICE INSTRUCTS THE USER TO; "VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE". THERE IS NO EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE (IFU0101-0). ROOT CAUSE REVIEW THE FAILURE OF NEEDLE BROKEN WAS CONCLUDED FROM THE AVAILABLE INFORMATION. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED FROM THE AVAILABLE INFORMATION. A POSSIBLE CAUSE COULD BE ATTRIBUTED TO DEVICE HANDLING OR EXCESSIVE FORCE BEING APPLIED CAUSING THE NEEDLE BREAKAGE WHEN ATTACHING OR DETACHING THE DEVICE TO THE SCOPE. SUMMARY COMPLAINT IS CONFIRMED BASED ON THE CUSTOMER'S TESTIMONY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.
AS REPORTED TO CUSTOMER RELATIONS, "THE DOCTOR WAS TO PERFORM A SOFT-TISSUE TRANSDUODENAL PUNCTURE WITHOUT THE USE OF THE ELEVATOR AND, WHEN PASSING THE NEEDLE ONCE, IT BROKE. ANOTHER NEEDLE WAS USED TO CONTINUE THE PROCEDURE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 41300 | ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE | FCG KIT, NEEDLE, BIOPSY | FCG | COOK IRELAND LTD | G31520 | C1457029 | 00827002315204 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |