FDA Adverse Event Other Summary report: N

CD1700 CS

MDR report key: 824500 · Received February 2, 2007

Report

Report Number
2919069-2007-00019
Event Type
Other
Date Received
February 2, 2007
Date of Event
January 8, 2007
Report Date
January 8, 2007
Manufacturer
ABBOTT DIAGNOSTICS DIVISION/CELLDYN
Product Code
JJE
PMA / PMN Number
k870233
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED NO INSTRUMENT FLAGS (DIFF FLAGS OR PLT FLAGS THAT WOULD REQUIRE SAMPLE VERIFICATION) BUT COULD NOT PROVIDE INFORMATION REGARDING DISPERSIONAL DATA ALERTS. THE CUSTOMER REPEATED RESULTS BASED ON THE LIS DELTA CHECK USING HISTORICAL DATA. A HGB VALUE OF 4.6 G/DL SHOULD HAVE GENERATED A DISPERSIONAL DATA ALERT. THE CD1700 SYSTEM OPERATOR'S MANUAL RECOMMENDS, ON PAGE 3-23, THAT RESULTS FALLING OUTSIDE ACTION LIMITS (DISPERSIONAL DATA ALERTS) BE CHECKED BY REPEATING THE SPECIMEN, PERFORMING A SMEAR REVIEW, OR NOTIFYING A PHYSICIAN. THE TROUBLESHOOTING SECTION ON PAGE 10-13 FOR ABNORMAL OR ERRATIC HGB, MCH, AND/OR MCHC HAS SUGGESTIONS WHICH INCLUDE: FLOW CELL CLEANING, RETESTING SAMPLE, CHECKING DIAGNOSTICS AND OBTAINING TECHNICAL ASSISTANCE. THE CUSTOMER STATED THAT DAILY MAINTENANCE WAS UP TO DATE. QC WAS ACCEPTABLE IN OPEN AND CLOSED MODES. THE CUSTOMER OBSERVED DURING AUTOCLEAN THAT THE TUBE HOLDER WELL LEAKED A LITTLE. THE CUSTOMER SUSPECTED A SHORT DRAW AS THE SOURCE OF THE ISSUE. TROUBLESHOOTING, PER THE DIRECTION OF THE CTA, FOUND NO INCOMPLETE ASPIRATION ERROR WAS GENERATED WHEN RUNNING THE CLOSED MODE WITH NO SAMPLE.THEREFORE, A FIELD SERVICE REPRESENTATIVE (FSR) WAS DISPATCHED TO CHECK THE INCOMPLETE ASPIRATION SENSOR. THE FSR FLUSHED OUT THE T-FITTING TUBE TO RESOLVE THE ISSUE. THE INSTRUMENT WAS VERIFIED BY RUNNING CONTROLS WHICH PASSED. IN ADDITION THE COMPLAINT ANALYSIS TRENDING SYSTEM (CATS) AND TRENDING ANALYSIS SUPPORT SYSTEM (TASS) REPORTS WERE REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED FOR THE ISSUE OF DISCREPANT RESULTS. THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

THE ACCOUNT STATED THAT A LOW HEMOGLOBIN (HGB) RESULT OF 4.6 G/DL WAS GENERATED IN THE CLOSED MODE ON THE CELLDYN 1700 ANALYZER. NO FLAGS WERE GENERATED. THE SAMPLE WAS REPEATED IN THE OPEN MODE AND THE HGB WAS 10.3 G/DL. THE LOW RESULT WAS NOT REPORTED AS IT WAS CONSIDERED SUSPECT. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CD1700 CS AUTOMATED HEMATOLOGY ANALYZER JJE ABBOTT DIAGNOSTICS DIVISION/CELLDYN NA NA

Patients

Seq Age Sex Outcome Treatment
1 YR