SOPRANO XL
Report
- Report Number
- 3004167969-2019-00001
- Event Type
- Injury
- Date Received
- January 11, 2019
- Date of Event
- August 10, 2018
- Report Date
- December 13, 2018
- Manufacturer
- ALMA LASERS LTD
- Product Code
- GEX
- UDI-DI
- 17290110120662
- PMA / PMN Number
- K083848
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- 501
Narratives
EXEMPTION LETTER E2013012 ALMA LTD. (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF ALMA, INC. (IMPORTER). SECTION H3: ALMA INC. (IMPORTER) HAS INSPECTED THE DEVICE AND IT APPEARS THAT THE UNIT MAY HAVE BEEN DROPPED OR HIT. THE OUTPUT WAS FOUND TO BE WITHIN THE MANUFACTURER'S SPECIFICATIONS BUT THE THERMOELECTRIC COUPLING (TEC) WAS NON-OPERATIONAL. ALMA LASERS MADE MULTIPLE ATTEMPTS TO REQUEST PATIENT PHOTOGRAPHS AND PERTINENT ADVERSE EVENT INFORMATION. THE FACILITY DID NOT PROVIDE THE ADVERSE EVENT DETAILS UNTIL THE DAY BEFORE THE REPORTABILITY 30-DAY DEADLINE. ALMA LASERS LTD. CLINICAL REVIEWED THE INFORMATION AND FOUND SOME OF THE PERTINENT INFORMATION TO BE LACKING. ALMA LASERS LTD. CLINICAL DETERMINED THAT VITAL INFORMATION ON PAST TREATMENT RECORDS NEED TO BE EXPLORED. IN THE ABSENCE OF VISUAL EVIDENCE AND PERTINENT CLINICAL INFORMATION, AT THIS TIME, NO ROOT CAUSE CAN BE ASSESSED. HOWEVER, IN GOOD FAITH EFFORTS, ALMA LASERS IS SUBMITTING THE REPORT TO THE FDA AT THIS TIME. ALMA LASERS IS INVESTIGATING THIS ISSUE AND WILL SUBMIT SUPPLEMENTAL REPORTS WITHIN THE FDA PUBLISHED TIMELINES, IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE SUSPECTED DEVICE WAS USED ON THE PATIENT'S UPPER BACK AND RIGHT LOWER ARM FOR HAIR REMOVAL. PATIENT SUSTAINED MULTIPLE AREAS OF ERYTHEMA ON THE UPPER BACK AND LEFT SIDE OF THE RIGHT ARM WITH BLISTERS. PATIENT ALSO SUSTAINED MULTIPLE BLISTERS ON THE BACK AND THE LEFT SIDE SHOULDER. THE FACILITY PROVIDED PATIENT WITH NEOSPORIN AND ADVISED AQUAPHOR TID.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 34094 | SOPRANO XL | MEDICAL LASER | GEX | ALMA LASERS LTD | 810 | 17290110120662 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR | Other |