FDA Adverse Event Malfunction Summary report: N

SMARTVIEW

MDR report key: 8239075 · Received January 11, 2019

Report

Report Number
1000165971-2019-00030
Event Type
Malfunction
Date Received
January 11, 2019
Date of Event
February 2, 2018
Report Date
October 3, 2019
Manufacturer
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
Product Code
LWP
PMA / PMN Number
UNKNOWN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

AN UPDATE WAS PROVIDED AS OF 17 MAY 2019: 93 HOME MONITORS WERE SUCCESSFULLY CONNECTED TO THE BACK-OFFICE; 5 PATIENTS WERE STILL PAIRED WITH A HOME MONITOR WITH AN ORANGE LIGHT.

Additional Manufacturer Narrative · 0

D4: PLEASE REFER TO THE DOCUMENT FILE-2019-00077_HM_CANADA_LISTSN_UPDATED.PDF FOR DEVICE SERIAL NUMBERS AFFECTED BY THIS EVENT. PLEASE REFER TO THE ATTACHED ANALYSIS REPORT. - ATTACHMENT: [FILE-2019-00077_HM_CANADA_LISTSN_UPDATED.PDF, 20190801 - FILE-2019-00077 - ANALYSIS_AND_CLOSURE_REPORT_RESP-2019-01065.PDF].

Additional Manufacturer Narrative · 0

THE LIST OF IMPACTED DEVICES HAS BEEN UPDATED, PLEASE REFER TO THE DOCUMENT FILE: (B)(4)CANADA_LISTSN_UPDATED. PDF FOR DEVICE SERIAL NUMBERS AFFECTED BY THIS EVENT. - ATTACHMENT: [FILE: (B)(4) CANADA_LISTSN_UPDATED.PDF].

Additional Manufacturer Narrative · 0

THE NUMBER OF UNITS IMPACTED BY THIS EVENT HAS INCREASED. PLEASE REFER TO THE DOCUMENT FILE-2019-00077_HM_CANADA_LISTSN_UPDATED.PDF FOR DEVICE SERIAL NUMBERS AFFECTED BY THIS EVENT. - ATTACHMENT: [FILE-2019-00077_HM_CANADA_LISTSN_UPDATED.PDF].

Additional Manufacturer Narrative · 0

AN UPDATE WAS PROVIDED AS OF 15 APRIL 2019: 66 HOME MONITORS WERE SUCCESSFULLY CONNECTED TO THE BACK-OFFICE; 11 PATIENTS WERE STILL PAIRED WITH A HOME MONITOR WITH AN ORANGE LIGHT.

Description of Event or Problem · 0

REPORTEDLY, AN ISSUE WAS OBSERVED TO SET UP NUMEROUS HOME MONITORS IN CANADA. AN ORANGE LIGHT INDICATOR ON THE MONITORS EVIDENCED DYSFUNCTION OF THE DEVICES. ONE OF THESE DEVICES WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 0

REPORTEDLY, AN ISSUE WAS OBSERVED TO SET UP NUMEROUS HOME MONITORS IN CANADA. AN ORANGE LIGHT INDICATOR ON THE MONITORS EVIDENCED DYSFUNCTION OF THE DEVICES.ONE OF THESE DEVICES WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 0

REPORTEDLY, AN ISSUE WAS OBSERVED TO SET UP NUMEROUS HOME MONITORS IN CANADA. AN ORANGE LIGHT INDICATOR ON THE MONITORS EVIDENCED DYSFUNCTION OF THE DEVICES. ONE OF THESE DEVICES WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 0

REPORTEDLY, AN ISSUE WAS OBSERVED TO SET UP NUMEROUS HOME MONITORS IN CANADA. AN ORANGE LIGHT INDICATOR ON THE MONITORS EVIDENCED DYSFUNCTION OF THE DEVICES.ONE OF THESE DEVICES WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 0

REPORTEDLY, AN ISSUE WAS OBSERVED TO SET UP NUMEROUS HOME MONITORS IN CANADA. AN ORANGE LIGHT INDICATOR ON THE MONITORS EVIDENCED DYSFUNCTION OF THE DEVICES.ONE OF THESE DEVICES WAS RETURNED FOR ANALYSIS.

Additional Manufacturer Narrative · 1

PLEASE REFER TO THE DOCUMENT FILE (B)(4) FOR DEVICE SERIAL NUMBERS AFFECTED BY THIS EVENT. (B)(4).

Description of Event or Problem · 1

REPORTEDLY, AN ISSUE WAS OBSERVED TO SET UP NUMEROUS HOME MONITORS IN (B)(6). AN ORANGE LIGHT INDICATOR ON THE MONITORS EVIDENCED DYSFUNCTION OF THE DEVICES. ONE OF THESE DEVICES WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
31484 SMARTVIEW PACEMAKER/ICD/CRT NON-IMPLANTED COMPONENTS LWP SORIN GROUP ITALIA S.R.L. - CRM FACILITY SMARTVIEW PSTN

Patients

Seq Age Sex Outcome Treatment
1