FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 8236416 · Received January 10, 2019

Report

Report Number
3005075853-2019-15673
Event Type
Malfunction
Date Received
January 10, 2019
Date of Event
December 19, 2018
Report Date
December 19, 2018
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
UDI-DI
10705036001843
PMA / PMN Number
K050344
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). BATCH # R94M14. DEVICE ANALYSIS: THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED WITH NO DAMAGE IN THE EXTERNAL COMPONENTS. IN ADDITION, 2 CLIPS PROPERLY FORMED WERE RETURNED INSIDE OF A PLASTIC BAG. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY. DURING THE ANALYSIS, THE DEVICE WAS CYCLED AND IT FED AND FORMED 5 CONFORMING CLIPS. UPON TESTING, THE JAWS OPEN AND CLOSE WITHOUT ANY DIFFICULTIES. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. THE BATCH HISTORY RECORD WAS REVIEWED AND NO DEFECTS, NCR¿S OR PROTOCOLS RELATED TO THE COMPLAINT, WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 0

(B)(4). DEVICE ANALYSIS: THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED WITH NO DAMAGE IN THE EXTERNAL COMPONENTS. IN ADDITION, 2 CLIPS WITH GAP NOT PROPERLY FORMED WERE RETURNED INSIDE OF A PLASTIC BAG. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY. DURING THE ANALYSIS, THE DEVICE WAS CYCLED AND IT FED AND FORMED 5 CONFORMING CLIPS. UPON TESTING, THE JAWS OPEN AND CLOSE WITHOUT ANY DIFFICULTIES. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED.

Additional Manufacturer Narrative · 1

(B)(4). BATCH # UNK. THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE DOCTOR FIRED TWO CLIPS ON TISSUE AND THE TWO CLIPS HAD FALLEN OFF THE TISSUE. THE DOCTOR RETRIEVED THE CLIPS FROM THE PATIENT AND USED A SECOND LIKE DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.