FDA Adverse Event Injury Summary report: N

TECNIS OPTIBLUE ITEC PRELOADED 1-PIECE IOL

MDR report key: 8234680 · Received January 10, 2019

Report

Report Number
2648035-2019-00053
Event Type
Injury
Date Received
January 10, 2019
Date of Event
December 11, 2018
Report Date
November 6, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
HQL
UDI-DI
05050474561762
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-(B)(4) AND CAPA-010215.

Additional Manufacturer Narrative · 1

IF IMPLANTED; GIVE DATE: N/A (NOT APPLICABLE). THE INTRAOCULAR LENS WAS REMOVED AND REPLACED DURING THE SAME PROCEDURE. IF EXPLANTED; GIVE DATE: N/A (NOT APPLICABLE). THE INTRAOCULAR LENS WAS REMOVED AND REPLACED DURING THE SAME PROCEDURE. (B)(6). THIS REPORT IS BEING FILED ON AN INTERNATIONAL DEVICE; TECNIS OPTIBLUE 1-PIECE IOL, MODEL PCB00V THAT HAS A SIMILAR DEVICE, TECNIS 1-PIECE IOL MODEL PCB00 WHICH IS DISTRIBUTED IN THE UNITES STATES UNDER PMA P980040. (B)(4). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE WAS NOT RETURNED AT THE MANUFACTURING SITE; THEREFORE, PRODUCT TESTING COULD NOT BE PERFORMED AND THE REPORTED COMPLAINT COULD NOT BE VERIFIED. MANUFACTURING RECORDS REVIEW: THE MANUFACTURING RECORDS FOR THE DEVICE WERE REVIEWED AND THERE WERE NO NON-CONFORMANCES WITH RESPECT TO THE MANUFACTURING PROCESS. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. HISTORICAL DATA ANALYSIS: A SEARCH IN THE COMPLAINTS HISTORY REVEALED THAT NO ADDITIONAL INVESTIGATION REQUEST HAS BEEN RECEIVED FOR THIS PRODUCTION ORDER NUMBER. LABELING REVIEW: THE DIRECTIONS FOR USE (DFU) WAS REVIEWED. THE DFU ADEQUATELY PROVIDES INSTRUCTIONS, PRECAUTIONS, AND WARNINGS FOR THE PROPER USE AND HANDLING OF THE PRODUCT. CONCLUSION: AS A RESULT OF THE INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT MODEL PCB00V MONOFOCAL IOL (INTRAOCULAR LENS) DID NOT COME OUT WHEN TURNING THE PLUNGER, SO THE PLUNGER WAS PUSHED IN. AS A RESULT, THE IOL POPPED OUT AND THEN CAPSULE BROKE. ANOTHER LENS WAS SUTURED INSTEAD AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THROUGH FOLLOW-UP, IT LEARNT THAT THE DOCTOR IS OBSERVING PATIENT'S PROGRESS AS THERE ARE NO HEALTH ISSUES SO FAR. ALSO THE REPLACEMENT LENS WAS A THREE PIECE LENS OF ANOTHER COMPANY. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27389 TECNIS OPTIBLUE ITEC PRELOADED 1-PIECE IOL MONOFOCAL IOLS HQL JOHNSON & JOHNSON SURGICAL VISION, INC. PCB00V 05050474561762

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention