TECNIS OPTIBLUE ITEC PRELOADED 1-PIECE IOL
Report
- Report Number
- 2648035-2019-00053
- Event Type
- Injury
- Date Received
- January 10, 2019
- Date of Event
- December 11, 2018
- Report Date
- November 6, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- HQL
- UDI-DI
- 05050474561762
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-(B)(4) AND CAPA-010215.
IF IMPLANTED; GIVE DATE: N/A (NOT APPLICABLE). THE INTRAOCULAR LENS WAS REMOVED AND REPLACED DURING THE SAME PROCEDURE. IF EXPLANTED; GIVE DATE: N/A (NOT APPLICABLE). THE INTRAOCULAR LENS WAS REMOVED AND REPLACED DURING THE SAME PROCEDURE. (B)(6). THIS REPORT IS BEING FILED ON AN INTERNATIONAL DEVICE; TECNIS OPTIBLUE 1-PIECE IOL, MODEL PCB00V THAT HAS A SIMILAR DEVICE, TECNIS 1-PIECE IOL MODEL PCB00 WHICH IS DISTRIBUTED IN THE UNITES STATES UNDER PMA P980040. (B)(4). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
DEVICE EVALUATION: THE DEVICE WAS NOT RETURNED AT THE MANUFACTURING SITE; THEREFORE, PRODUCT TESTING COULD NOT BE PERFORMED AND THE REPORTED COMPLAINT COULD NOT BE VERIFIED. MANUFACTURING RECORDS REVIEW: THE MANUFACTURING RECORDS FOR THE DEVICE WERE REVIEWED AND THERE WERE NO NON-CONFORMANCES WITH RESPECT TO THE MANUFACTURING PROCESS. THE PRODUCT WAS MANUFACTURED AND RELEASED ACCORDING TO SPECIFICATION. HISTORICAL DATA ANALYSIS: A SEARCH IN THE COMPLAINTS HISTORY REVEALED THAT NO ADDITIONAL INVESTIGATION REQUEST HAS BEEN RECEIVED FOR THIS PRODUCTION ORDER NUMBER. LABELING REVIEW: THE DIRECTIONS FOR USE (DFU) WAS REVIEWED. THE DFU ADEQUATELY PROVIDES INSTRUCTIONS, PRECAUTIONS, AND WARNINGS FOR THE PROPER USE AND HANDLING OF THE PRODUCT. CONCLUSION: AS A RESULT OF THE INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON AND JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT MODEL PCB00V MONOFOCAL IOL (INTRAOCULAR LENS) DID NOT COME OUT WHEN TURNING THE PLUNGER, SO THE PLUNGER WAS PUSHED IN. AS A RESULT, THE IOL POPPED OUT AND THEN CAPSULE BROKE. ANOTHER LENS WAS SUTURED INSTEAD AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THROUGH FOLLOW-UP, IT LEARNT THAT THE DOCTOR IS OBSERVING PATIENT'S PROGRESS AS THERE ARE NO HEALTH ISSUES SO FAR. ALSO THE REPLACEMENT LENS WAS A THREE PIECE LENS OF ANOTHER COMPANY. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 27389 | TECNIS OPTIBLUE ITEC PRELOADED 1-PIECE IOL | MONOFOCAL IOLS | HQL | JOHNSON & JOHNSON SURGICAL VISION, INC. | PCB00V | 05050474561762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |