BD¿ SYRINGE WITH BD LUER-LOK¿ TIP
Report
- Report Number
- 1911916-2018-00832
- Event Type
- Malfunction
- Date Received
- January 10, 2019
- Date of Event
- December 24, 2018
- Report Date
- February 19, 2019
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- UDI-DI
- 30382903096535
- PMA / PMN Number
- K980987
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: ONE SAMPLE WAS RECEIVED FOR EVALUATION. IT WAS TESTED FOR VOLUMETRIC ACCURACY. GIVING 10.16; THE SPECIFICATION IS 9.5 GRAMS TO 10.50 GRAMS. IT IS WITHIN SPECIFICATION THEREFORE FAILURE MODE IS NOT VERIFIED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. THIS IS THE 1ST COMPLAINT FOR THE LOT# 6252555 FOR THE SAME DEFECTS OR SYMPTOMS. THERE WAS NO DOCUMENTATION OF ISSUES FOR THE COMPLAINT OF LOT # 6252555 DURING THIS PRODUCTION RUN. FAILURE NOT CONFIRMED.
IT WAS REPORTED WITH THE USE OF THE BD¿ SYRINGE WITH BD LUER-LOK¿ TIP THERE WAS AN ISSUE WITH READING 52 ML WHEN 50 ML OF LIQUID WAS DRAWN.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WITH THE USE OF THE BD¿ SYRINGE WITH BD LUER-LOK¿ TIP THERE WAS AN ISSUE WITH READING 52 ML WHEN 50 ML OF LIQUID WAS DRAWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 25921 | BD¿ SYRINGE WITH BD LUER-LOK¿ TIP | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 6252555 | 30382903096535 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |