FDA Adverse Event Malfunction Summary report: N

BD¿ SYRINGE TIP CAP

MDR report key: 8233225 · Received January 10, 2019

Report

Report Number
1911916-2018-00834
Event Type
Malfunction
Date Received
January 10, 2019
Date of Event
December 21, 2018
Report Date
February 21, 2019
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
KYW
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: TEN REPRESENTATIVE SAMPLES WERE RECEIVED. THEY WERE VISUALLY INSPECTED UNDER THE MICROSCOPE. A FUNCTIONAL TEST WAS PERFORMED BY CONNECTING THEM TO A SALINE SYRINGE, THE PLUNGER ROD OF THE SYRINGE WAS PUSHED DOWN TRYING TO EXPEL THE SALINE SOLUTION. NO LEAKAGE WAS OBSERVED, THEREFORE FAILURE MODE IS NOT VERIFIED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES.

Description of Event or Problem · 0

IT WAS REPORTED WITH THE USE OF THE BD¿ SYRINGE TIP CAP THERE WERE 2 ISSUES WITH LEAKAGE FROM THE SEAL BETWEEN THE CAP AND THE SYRINGE, ONCE WITH LOT# 7200661 AND ONCE WITH 8207911.

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 7200661. MEDICAL DEVICE EXPIRATION DATE: 2022-09-30. DEVICE MANUFACTURE DATE: 2017-07-19. MEDICAL DEVICE LOT #: 8207911. MEDICAL DEVICE EXPIRATION DATE: 2023-06-30. DEVICE MANUFACTURE DATE: 2018-07-26. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WITH THE USE OF THE BD¿ SYRINGE TIP CAP THERE WERE 2 ISSUES WITH LEAKAGE FROM THE SEAL BETWEEN THE CAP AND THE SYRINGE, ONCE WITH LOT# 7200661 AND ONCE WITH 8207911.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27813 BD¿ SYRINGE TIP CAP LIQUID MEDICATION DISPENSER KYW BD MEDICAL (BD WEST) MEDICAL SURGICAL SEE H.10

Patients

Seq Age Sex Outcome Treatment
1 Other