FDA Adverse Event Injury Summary report: N

HME BOOSTER

MDR report key: 823152 · Received March 2, 2007

Report

Report Number
1824226-2007-00001
Event Type
Injury
Date Received
March 2, 2007
Date of Event
January 25, 2007
Report Date
February 6, 2007
Manufacturer
KING SYSTEMS CORP.
Product Code
BTT
PMA / PMN Number
K990614
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: KING SYSTEMS AND MEDISIZES HAVE DETERMINED THAT THIS TYPE OF EVENT CAN HAPPEN IF WATER LEAKS INTO THE BACKSIDE OF THE HEATER ELEMENT. THIS LEAKAGE CAN OCCUR IF THE USER DOES NOT FOLLOW THE INSTRUCTIONS FOR USE, WHICH IS INCLUDED WITH THE DEVICE. SPECIALLY: TURN OFF THE WATER FEED SET BEFORE REMOVING THE HEATING ELEMENT FROM THE T-PIECE. ALWAYS USE THE FLOW REGULATOR TO LIMIT THE WATER AVAILABLE TO 10 ML PER HOUR. WHEN IN USE, POSITION THE HEATER CORD BELOW THE HORIZONTAL PLANE. POSITION THE WATER FEED SET SO THAT THE LINE DOES NOT CROSS THE ELECTRICAL COMPONENTS. CHECK IF THE O-RING ON THE HEATER IS IN GOOD CONDITION. NEVER IMMERSE THE HME-BOOSTER HEATER IN LIQUID.

Description of Event or Problem · 1

THE USER-FACILITIE REPORTED: PT HAS RECEIVED A BURN ON HIS LEFT FOREARM WE THINK THE BURN CAME FROM A FAULTY HME BOOSTER THAT WAS INLINE WITH HIS VENTILATOR CIRCUIT. PT HAD BLISTERS ON HIS LEFT FOREARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HME BOOSTER HEATER BTT KING SYSTEMS CORP. MSB-600 *

Patients

Seq Age Sex Outcome Treatment
1 * Required Intervention