FDA Adverse Event Malfunction Summary report: N

MPACT IMPACTION PLATE DML

MDR report key: 8230884 · Received January 9, 2019

Report

Report Number
3005180920-2018-01077
Event Type
Malfunction
Date Received
January 9, 2019
Date of Event
December 10, 2018
Report Date
January 9, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 08 JANUARY 2019: LOT 1654048: 21 ITEMS MANUFACTURED AND RELEASED ON 17-MAY-2017. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, NO SIMILAR EVENT HAS BEEN REPORTED ON THIS LOT. PRELIMINARY INVESTIGATION PERFORMED ON 08 JAN 2019 BY R&D PRODUCT MANAGER: THE PHOTOS SHOW THE PIECE WITH THE LOCKING COMPONENT SEPARATED FORM THE PLATE. THE EVENT CAN BE RELATED TO AN HIGH TORSION FORCE THAT COULD HAVE PROVIDED THE BREAKAGE OF THE LOCKING COMPONENT.

Description of Event or Problem · 1

DURING THE PRIMARY HIP SURGERY AND AFTER IMPACTING THE CUP, THE IMPACTION PLATE BROKE AND A FRAGMENT FELL INTO THE PATIENT AND WAS REMOVED. THERE WAS NO DELAY IN THE CASE AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
23594 MPACT IMPACTION PLATE DML SURGICAL INSTRUMENT FOR HIP LXH MEDACTA INTERNATIONAL SA 1654048

Patients

Seq Age Sex Outcome Treatment
1 Other