FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 8228679 · Received January 8, 2019

Report

Report Number
1645337-2019-07832
Event Type
Injury
Date Received
January 8, 2019
Date of Event
January 27, 2018
Report Date
December 12, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001287
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2019, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. DEVICE EVALUATION SUMMARY: THE DEVICE WAS RETURNED TO MENTOR WITHOUT FLUID INSIDE. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. MENTOR PRODUCT ANALYSIS LAB DISCOVERED A RENT MEASURING APPROXIMATELY 0.1 CM NEAR THE VALVE. MICROSCOPIC EXAMINATION WAS PERFORMED. NO EVIDENCE OF INSTRUMENT DAMAGE WAS OBSERVED. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE, NOR THE ETIOLOGY OF THE YELLOW MATERIAL FOUND ON THE DEVICE. A MANUFACTURING RECORD EVALUATION (MRE) OF LOT NUMBER 202698 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED TO MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES, AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT MEDICAL PRODUCTS: MENTOR SMOOTH ROUND MODERATE PROFILE 425CC, CATALOG NUMBER 3501670, LOT NUMBER 202698. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION PRIMARY WITH A MENTOR SMOOTH ROUND MODERATE PROFILE 425CC BREAST IMPLANT AND EXPERIENCED DEFLATION ON THE RIGHT SIDE. AS A RESULT, THE PATIENT UNDERWENT REMOVAL ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
22076 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 202698 00081317001287

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention SEE H10