FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS

MDR report key: 82282 · Received February 20, 1997

Report

Report Number
1527736-1997-00071
Event Type
Malfunction
Date Received
February 20, 1997
Date of Event
January 24, 1997
Report Date
February 20, 1997
Manufacturer
EES-CINCINNATI
Product Code
KOG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

D6: DEVICE RETURNED WITH NO LOT# IDENTIFICATION. BASED UPON THE INQUIRY INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE RETURNED CARTRIDGE HAD DAMAGETO THE LOCKOUT TABS WHICH INDICATES THE FIRING CYCLE WAS STOPPED, RETURNED, THEN STARTED. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE INSTRUMENT WAS CYCLED, FIRED, CUT AND FORMED THE STAPLES WITHIN DESIGN SPECIFICATION. THE INSTRUMENT WAS DISASSEMBLED TO EXAMINE THE INTERNAL COMPONENTS AND NO DEFORMATION COULD BE INDENTIFIED. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECIFICATIONS. IF THE INSTRUMENTS'S FIRING CYCLE IS STOPPED, RETURNED, THEN RESTARTED, THE INSTRUMENT AND CARTRIDGE MAY BECOME DAMAGED. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE CO'S PRODUCTS.

Description of Event or Problem · 1

DURING A LAPAROSCOPIC ASSISTED VAGINAL HYSTERECTOMY THE TSW35 INADVERTENTLY LOCKED OUT ON THE SECOND FIRING. THE INSTRUMENT WAS RELOADED AND WORKED FINE. IT WAS USED FOR A TOTAL OF FIVE FIRINGS. THE RELOAD OUT WILL BE RETURNED. THERE WAS NO CONSEQUENCE TO THE PT. CLINICAL FOLLOW-UP 1/27/97 1540 OFFICE CLOSED FOR LUNCH. 1/27/97 1645 THE SURGEON STATED THE DEVICE WOULD NOT FIRE ON THE SECOND FIRING. HE STATED THE NURSE HAD NOT SQUEEZED THE DEVICE. THE STAPLES WERE COMING OUT UNFORMED FROM THE DEVICE. THE DEVICE WAS FIRED SEVERAL ADD'L TIMES WITHOUT DIFFICULTY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS LINEAR CUTTER KOG EES-CINCINNATI NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other