FDA Adverse Event Malfunction Summary report: N

SLALOM THRILL PTA BALLOON CATHETER

MDR report key: 822805 · Received January 25, 2007

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
9610978-2007-00059
Event Type
Malfunction
Date Received
January 25, 2007
Date of Event
December 28, 2006
Report Date
December 29, 2006
Manufacturer
CORDIS EUROPA, N.V.
Product Code
LTI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PLEASE NOTE THAT THE SLALOM THRILL PTA CATHETER IS SIMILAR TO THE SLALOM PTA CATHETER. THE PRODUCT WAS NOT RETURNED FOR ANALYSIS AND EVALUATION. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED AND SHOWED THAT THIS LOT OF PRODUCTS (R0406428) MET ALL REQUIREMENTS PER THE APPLICABLE MANUFACTURING QUALITY PLAN (MQP). THIS REVIEW WAS EXTENDED TO SUBASSEMBLY PART NUMBER E7144631 WITH LOT NUMBER 0304060223. THIS REVIEW CONFIRMED THAT SUBASSEMBLIES MET ALL REQUIREMENTS PER THE APPLICABLE MQP AS WELL, INCLUDING THE BURST TEST VALUES. WITH THE LIMITED AMOUNT OF INFO AVAILABLE AND WITHOUT THE RETURN OF THE PRODUCT OR FILMS OF THE PROCEDURE, IT IS NOT POSSIBLE TO DETERMINE THE RELATIONSHIP BETWEEN THE DEVICE AND THE EVENT. HOWEVER, VESSEL CHARACTERISTICS MAY HAVE CONTRIBUTED TO THE REPORT EVENT.

Description of Event or Problem · 1

DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TO THE LEFT SHUNT ANASTOMOSIS, THE BALLOON WAS REPORTED TO HAVE RUPTURED. THE VESSEL HAD SEVERE CALCIFICATION, SEVERE TORTUOSITY AND A 90% STENOSIS. A GUIDEWIRE WAS INSERTED ACROSS THE LESION VIA SHEATH INTRODUCER. THE BALLOON CATHETER WAS ADVANCED TO THE LESION OVER THE GUIDEWIRE THROUGH THE SHEATH INTRODUCER AND THE BALLOON WAS INFLATED USING AN INDEFLATOR. HOWEVER, THE BALLOON RUPTURED AT BELOW NOMINAL PRESSURE DURING THE FIRST INFLATION. THE BALLOON CATHETER WAS WITHDRAWN FROM THE PT'S BODY AND ANOTHER BALLOON CATHETER WAS USED TO END PROCEDURE SUCCESSFULLY. THE PRODUCT WAS PREPARED AND USED AS PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO REPORTED PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SLALOM THRILL PTA BALLOON CATHETER PTA CATHETER (LIT) LTI CORDIS EUROPA, N.V. NA R0406428

Patients

Seq Age Sex Outcome Treatment
1 * SLALOM THRILL/CORDIS (439-6020T)| 6F SHEATH INTRODUCER/MEDIKIT| 0.016 INCH/TERUMO| GT WIRE| INDEFLATOR ENCORE/BOSTON