FDA Adverse Event Malfunction Summary report: Y

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 8226591 · Received January 8, 2019

Report

Report Number
2531779-2019-00171
Event Type
Malfunction
Date Received
January 8, 2019
Report Date
January 1, 2019
Manufacturer
ANIMAS LLC
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

OF THE 25 ALLEGATIONS OF A MALFUNCTION, 7 PUMPS WERE RETURNED TO ANIMAS AND INVESTIGATED. OF THE INVESTIGATED PUMPS, WERE FOUND TO BE MALFUNCTIONING DUE TO TRANSCEIVER CONNECTOR FAILURES. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: IN QUARTER 1 OF 2019, THERE WERE 3 INSTANCES OF CS 052/053/054/055 SLEEP FAILURE FOUND DURING INVESTIGATION. THIS REPORT IS PROCESSED UNDER THE VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM.

Description of Event or Problem · 1

THIS REPORT SUMMARIZES <NOE> 25</NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THE ONE TOUCH PING MODEL PUMP EXPERIENCED A CS 052/053/054/055 SLEEP ISSUE. THESE REPORTS WERE RECEIVED FROM VARIOUS SOURCES. THE PATIENTS ASSOCIATED WITH THIS ALLEGATION RANGED FROM 4-70 YEARS OF AGE AND BETWEEN 38 AND 260 POUNDS. OF THE REPORTED PATIENTS, 14 WERE MALE AND 11 WERE FEMALE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
19903 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS LLC

Patients

Seq Age Sex Outcome Treatment
1