91-SERIES
Report
- Report Number
- 9616031-2019-00001
- Event Type
- Malfunction
- Date Received
- January 8, 2019
- Report Date
- February 4, 2019
- Manufacturer
- GETINGE DISINFECTION AB
- Product Code
- MEC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS REPORT IS BEING FILED UNDER EXEMPTION E2017050 BY THE MANUFACTURER GETINGE DISINFECTION AB, (REGISTRATION NO. 9616031) ON BEHALF OF THE IMPORTER GETINGE GROUP LOGISTICS AMERICA, LLC (REGISTRATION NO. 3012092534). WE ARE AWARE THAT THIS IS PAST THE 30 DAY DEADLINE FOR REPORTING. WE HAVE REMINDED THE COMPLAINT HANDLER TO REVIEW FDA REPORTING GUIDELINES IN ORDER TO PREVENT THIS FROM HAPPENING AGAIN. THE ISSUE IS BEING INVESTIGATED BY MANUFACTURING SITE.
THIS REPORT IS BEING FILED UNDER EXEMPTION E2017050 BY THE MANUFACTURER GETINGE DISINFECTION AB, (REGISTRATION NO. 9616031) ON BEHALF OF THE IMPORTER GETINGE GROUP LOGISTICS AMERICA, LLC (REGISTRATION NO. 3012092534). WHEN REVIEWING REPORTABLE EVENTS FOR THE DEVICES BELONGING TO 91 SERIES, WE WERE ABLE TO ESTABLISH THAT THIS ISSUE IS A FIRST ONE WHERE A STEAM LEAK FROM THE DOOR AREA WAS REPORTED TO US. THE DEVICE INVOLVED IN THE INCIDENT IS A WASHER- DISINFECTOR, MODEL NUMBER 9120, WITH THE SERIAL NUMBER (B)(4). . THE UNIT WAS MANUFACTURED ON (B)(6) 2007 AND INSTALLED ON (B)(6) 2007. AS SUCH IT HAS BEEN IN USE FOR OVER 11 YEARS. BASED ON THE INFORMATION COLLECTED TO DATE WE CONFIRMED THAT AN ALARM OCCURRED DUE TO CLOGGED STEAM LINE. ACCORDING TO SERVICE MANUAL (B)(4). REV. B (DOC. REF. NO.), IN SUCH AN ALARM SITUATION, THE TECHNICIAN OR TRAINED OPERATOR OF THE DEVICE IS TO PRESS THE "CLEAR ALARM" BUTTON ON THE OPERATING PANEL THAT INDICATES ABORTING THE CURRENT PROCESS. OPERATOR IS TO ACKNOWLEDGE AN ERROR BY ENTERING PASSWORD TO THE CONTROL PANEL DISPLAY. AFTERWARDS, THE DEVICE IS TO BE DRAINED MANUALLY WITH CAUTION AND THE DOOR IS NOT TO BE OPENED IF AN INTERNAL TEMPERATURE IS HIGHER THAN 60 DEGREES. IF AN EMERGENCY OCCURS, THE MACHINE IS TO BE STOPPED WITH ONE OF THE EMERGENCY STOPS. IN THE INVESTIGATED SITUATION, IT APPEARS MOST LIKELY THAT AFTER THE ERROR WAS CLEARED FROM THE MACHINE IN ACCORDANCE TO GIVEN INSTRUCTION, THE USER OPENED THE DOOR OF THE DEVICE WITHOUT CHECKING THE TEMPERATURE INDICATION FIRST. THIS IMMEDIATE ACTION OF OPENING THE DOOR WITHOUT FOLLOWING THE INSTRUCTIONS TO NOT TO DO SO AFTER CHECKING THE TEMPERATURE AND WAITING FOR THE INTERNAL TEMPERATURE TO COOL DOWN RESULTED IN AN EXTERNAL LEAKAGE OF THE STEAM AND, CONSEQUENTLY, AN ACTIVATION OF THE FIRE ALARM. THERE WAS NO PERSONAL INJURY AS A RESULT OF THE STEAM ESCAPING FROM THE DEVICE, NOR FROM ANY REACTION TO THE FIRE ALARM. THIS ACTIVITY WAS POSSIBLE DUE TO THE WAY DEVICES OF THIS AGE AND BUILD YEAR WERE DESIGNED TO WORK. IT IS CLEAR FROM THESE FINDINGS THAT THE ISSUE WAS CAUSED BY THE USER NOT FOLLOWING USE INSTRUCTIONS. IN SUMMARY, WHEN THE EVENT OCCURRED, THE DEVICE DID NOT MEET ITS SPECIFICATION AS A CLOGGED STEAM LINE CAUSED AN ALARM OCCURRENCE. THIS TYPE OF ISSUE IS CONSIDERED TO BE EXPECTED AND RELATED WITH A NORMAL USAGE OF THE DEVICE AND CAN BE AVOIDED BY FOLLOWING THE PREVENTIVE MAINTENANCE GIVEN IN THE MANUAL. THE REPORTED INCIDENT HOWEVER RESULTED FROM USER NOT FOLLOWING USE INSTRUCTIONS AND THIS, IN COMBINATION, CONTRIBUTED TO THE EVENT. IN THE TIME WHEN THE EVENT OCCURRED, THE DEVICE WAS NOT BEING USED FOR PATIENT TREATMENT. GIVEN THE FINDINGS OF THIS INVESTIGATION WE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DO NOT PROPOSE ANY FURTHER ACTION AT THIS TIME.
ON (B)(6) 2018 WE BECAME AWARE OF AN INCIDENT WITH WASHER DISINFECTOR 9120. AS IT WAS STATED, THE STEAM LINE WAS CLOGGED AND STEAM LEAK OCCURRED. THE STEAM FILLED IN THE WHOLE ROOM AND THE OPERATOR OF THE DEVICE WAS STANDING NEXT TO THE WASHER DISINFECTOR. THERE WAS NO INJURY REPORTED HOWEVER WE DECIDED TO REPORT THE ISSUE BASED ON THE POTENTIAL AS ANY UNEXPECTED STEAM LEAK FROM PARTS AVAILABLE FOR CUSTOMER MAY LEAD TO ADVERSE OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 18353 | 91-SERIES | 9120 | MEC | GETINGE DISINFECTION AB | 9120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |