ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAGETY LOCK
Report
- Report Number
- 1527736-1997-00065
- Event Type
- Malfunction
- Date Received
- February 20, 1997
- Date of Event
- January 22, 1997
- Report Date
- February 20, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
H6; CODE 400: DAMAGED FIRING MECHANISM. BASED UPON THE INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE INSTRUMENT WAS RETURNED NON-FUNCTIONAL. THE RETURNED CARTRIDGE HAD BROKEN ALIGNMENT FANGS. THE INSTRUMENT WAS DISASSEMBLED AND FOUND TO HAVE A DAMAGED FIRING MECHANSIM. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. SOME CONDITONS WHICH MIGHT RESULT IN DAMAGE TO THE FIRING MECHANISM ARE: INTERRUPTED FIRING CYLCE, TISSUE THICKER THAN INDICATED, ATTEMPTING TO FIRE THROUGH A SPENT CARTRIDGE, FIRING PRIOR TO COMPLETE CLAMPING OF THE JAWS AND FAILURE TO PROPERLY FOLLOW RELOADING INSTRUCTIONS. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMRPOVE CO'S PRODUCTS.
DURING A THORACOTOMY THE SURGEON WAS INSERTING THE EZ45G INTO THE CHEST WALL AND THE INSTRUMENT WOULD NOT OPEN. THE INSTRUMENT WAS REMOVED AND A SECOND INSTRUMENT WAS OPENED TO COMPLETE CASE. NO CONSEQUENCE TO THE PT. (NO BUTTRESSING MATERIAL WAS USED). 1/27/97 NO ANSWER. 1/27/97 1510 ATTEMPTED TO CONTACT SURGEON BUT NO ANSWER. 1/29/97 1331 NO ANSWER. 1/29/97 1333 CALLED DIRECTORY ASSISTANCE AND OBTAINED CORRECT PHONE NUMBER FOR SURGEON. 1/29/97 1334 MESSAGE AND 800# LEFT FOR MD TO CALL BACK. 1/29/97 1404 SURGEON CALLED BACK AND STATED ON APPROX THE THIRD FIRING, THE INSTRUMENT "WOULD NOT COME APART." IT WAS OPENED BY PRYING JAWS OPEN AND IN THE PROCESS THE HANDLE BROKE. IT DID NOT LAY STAPLES OR CUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAGETY LOCK | ENDOSCOPIC LINEAR CUTTER | GAG | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |