FDA Adverse Event Malfunction Summary report: N

ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAGETY LOCK

MDR report key: 82259 · Received February 20, 1997

Report

Report Number
1527736-1997-00065
Event Type
Malfunction
Date Received
February 20, 1997
Date of Event
January 22, 1997
Report Date
February 20, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GAG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

H6; CODE 400: DAMAGED FIRING MECHANISM. BASED UPON THE INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE INSTRUMENT WAS RETURNED NON-FUNCTIONAL. THE RETURNED CARTRIDGE HAD BROKEN ALIGNMENT FANGS. THE INSTRUMENT WAS DISASSEMBLED AND FOUND TO HAVE A DAMAGED FIRING MECHANSIM. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. SOME CONDITONS WHICH MIGHT RESULT IN DAMAGE TO THE FIRING MECHANISM ARE: INTERRUPTED FIRING CYLCE, TISSUE THICKER THAN INDICATED, ATTEMPTING TO FIRE THROUGH A SPENT CARTRIDGE, FIRING PRIOR TO COMPLETE CLAMPING OF THE JAWS AND FAILURE TO PROPERLY FOLLOW RELOADING INSTRUCTIONS. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMRPOVE CO'S PRODUCTS.

Description of Event or Problem · 1

DURING A THORACOTOMY THE SURGEON WAS INSERTING THE EZ45G INTO THE CHEST WALL AND THE INSTRUMENT WOULD NOT OPEN. THE INSTRUMENT WAS REMOVED AND A SECOND INSTRUMENT WAS OPENED TO COMPLETE CASE. NO CONSEQUENCE TO THE PT. (NO BUTTRESSING MATERIAL WAS USED). 1/27/97 NO ANSWER. 1/27/97 1510 ATTEMPTED TO CONTACT SURGEON BUT NO ANSWER. 1/29/97 1331 NO ANSWER. 1/29/97 1333 CALLED DIRECTORY ASSISTANCE AND OBTAINED CORRECT PHONE NUMBER FOR SURGEON. 1/29/97 1334 MESSAGE AND 800# LEFT FOR MD TO CALL BACK. 1/29/97 1404 SURGEON CALLED BACK AND STATED ON APPROX THE THIRD FIRING, THE INSTRUMENT "WOULD NOT COME APART." IT WAS OPENED BY PRYING JAWS OPEN AND IN THE PROCESS THE HANDLE BROKE. IT DID NOT LAY STAPLES OR CUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH THORACIC ENDO LINEAR CUTTER W/SAGETY LOCK ENDOSCOPIC LINEAR CUTTER GAG ETHICON ENDO-SURGERY, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other