FDA Adverse Event Malfunction Summary report: N

EVOLIS MICROPLATE PROCESSOR SYSTEM

MDR report key: 8224339 · Received January 7, 2019

Report

Report Number
3022521-2018-00008
Event Type
Malfunction
Date Received
January 7, 2019
Date of Event
December 13, 2018
Report Date
January 7, 2019
Manufacturer
BIO-RAD LABORATORIES
Product Code
JJE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

BIO-RAD TECHNICAL SUPPORT WAS CONTACTED BY (B)(6) ON (B)(6) 2018, TO REPORT THAT THE POWER STRIP CONNECTED TO THE EVOLIS MICROPLATE SYSTEM SPARKED WHEN THE POWER WAS TURNED ON. A BIO-RAD LABORATORIES FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE FACILITY TO INVESTIGATE THE INCIDENT. THE FSE DISCOVERED THAT A NYLON RETAINING NUT THAT SECURES THE LIQUID WASTE TUBING INTO THE LIQUID WASTE CONTAINER HAD COME OFF WHICH ALLOWED THE LIQUID WASTE TUBING TO COME OUT OF THE LIQUID WASTE CONTAINER. THE LIQUID WASTE WAS PUMPED INTO THE SYSTEM CABINET WHERE THE POWER STRIP AND UNINTERRUPTIBLE POWER SUPPLY (UPS) ARE LOCATED. SOME OF THE LIQUID WASTE WAS ABLE TO GET ON/IN THE POWER STRIP AND WHEN POWER WAS APPLIED THIS CAUSED THE POWER STRIP TO SPARK. ADDITIONALLY, THE SPARK LEFT A SCORCH MARK ON THE POWER STRIP. THE FSE REPLACED THE POWER STRIP, UPS, AND LIQUID WASTE CONTAINER (WHICH INCLUDES THE NYLON RETAINING NUT). ALSO AS PART OF THE INVESTIGATION THE FSE PERFORMED PERFORMANCE VERIFICATION TESTING TO ENSURE THE SYSTEM MET ALL MANUFACTURER SPECIFICATIONS. ALL TESTS MET MANUFACTURER SPECIFICATION AND THE SYSTEM WAS RETURN TO FULL OPERATION PRIOR TO RELEASING THE SYSTEM TO THE CUSTOMER. BIO-RAD LABORATORIES IS REPORTING THIS ISSUE AS AN MDR DUE TO THE FOLLOWING RISKS EVEN THOUGH THERE WAS NO HARM TO PATIENT OR END USER DURING THIS INCIDENT: SINCE THE POWER STRIP SPARKED IT WAS A POTENTIAL FIRE HAZARD, AND THERE WAS INCREASED RISK TO END USER FOR EXPOSURE TO BIO-HAZARDOUS WASTE SINCE THE LIQUID WASTE WAS NOT BEING PUMPED INTO THE PROPER CONTAINER. BIO-RAD LABORATORIES HAS REVISED THE EVOLIS PREVENTIVE MAINTENANCE REPORT TO INCLUDE INSPECTION OF THE NYLON RETENTION NUT FOR THE LIQUID WASTE TUBING TO REDUCE THE RISK OF RECURRENCE OF THIS INCIDENT. THE CHANGE IS LISTED UNDER THE SIMI-ANNUAL PM KIT PARTS SECTION SPECIFIC TO THE EXTERNAL LIQUID WASTE TUBING, AND STATES: "ALSO VERIFY THE WASTE BOTTLE TUBING KEEPER NUT IS MOUNTED PROPERLY AND IN GOOD CONDITION. REPLACE AS NEEDED 14MM PLASTIC NUT OR ENTIRE BOTTLE PN 89715." EVOLIS PREVENTIVE MAINTENANCE REPORT. FIELD SERVICE PERSONNEL WILL BE NOTIFIED OF THE CHANGE TO THE PREVENTIVE MAINTENANCE REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
15832 EVOLIS MICROPLATE PROCESSOR SYSTEM EVOLIS JJE BIO-RAD LABORATORIES

Patients

Seq Age Sex Outcome Treatment
1