ENDOPATH ETS
Report
- Report Number
- 1527736-1997-00167
- Event Type
- Malfunction
- Date Received
- February 19, 1997
- Date of Event
- January 22, 1997
- Report Date
- February 19, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- KOG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
H6; CODE 400: CARTRIDE LOCKOUT DAMAGED, AS IF REFIRED AFTER FIRING. BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE INSTRUMENT REPORTEDLY "MISFIRED" DURING SURGERY. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE INSTRUMENT WAS CYCLED, FIRED, CUT, AND FORMED THE STAPLES WITHIN DESIGN SPECIFICATION. THE INSTRUMENT WAS DISASSEMBLED TO EXAMINE THE INTERNAL COMPONENTS AND NO DEFORMATIONS COUD BE IDENTIFIED. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECIFICATIONS. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE IT FUNCTIONS PROPERLY. CO STIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE CO'S PRODUCTS.
DURING A LAPAROSCOPIC ASSISTED VAGINAL HYSTERECTOMY, THE SURGEON FIRED THE TSW35 SIX TIMES WITHOUT DIFFICULTY. ON THE SEVENTH FIRING THE DEVICE MISFIRED. A SECOND DEVICE WAS OPENED TO COMPLETE THE CASE. THERE WAS NO CONSEQUENCE TO THE PT. CLINICAL FOLLOW-UP: 1/24/97 1340 SURGEON DESCRIBED SEVERAL DIFFICULTIES WITH THE INSTRUMENT: SOMETIMES THE INSTRUMENT "UNLOCKED BEFORE FIRING," SOMETIMES IT WOULD NOT OPEN AFTER FIRING, AND IT FIRED STAPLES BUT DID NOT CUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ETS | LINEAR CUTTER | KOG | ETHICON ENDO-SURGERY, INC. | NA | J4500F |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |