FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS

MDR report key: 82239 · Received February 19, 1997

Report

Report Number
1527736-1997-00167
Event Type
Malfunction
Date Received
February 19, 1997
Date of Event
January 22, 1997
Report Date
February 19, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
KOG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

H6; CODE 400: CARTRIDE LOCKOUT DAMAGED, AS IF REFIRED AFTER FIRING. BASED UPON THE INQUIRY INFO RECEIVED, THE VISUAL EXAMINATION, AND THE FUNCTIONAL TESTING, NO CONCLUSION COULD BE REACHED AS TO WHY THE INSTRUMENT REPORTEDLY "MISFIRED" DURING SURGERY. THE INSTRUMENT WAS RETURNED IN GOOD PHYSICAL CONDITION. THE INSTRUMENT WAS CYCLED, FIRED, CUT, AND FORMED THE STAPLES WITHIN DESIGN SPECIFICATION. THE INSTRUMENT WAS DISASSEMBLED TO EXAMINE THE INTERNAL COMPONENTS AND NO DEFORMATIONS COUD BE IDENTIFIED. IT WAS CONCLUDED THAT THE INSTRUMENT WAS FULLY FUNCTIONAL AND CONFORMING TO DESIGN SPECIFICATIONS. THE EXPERIENCE THE SURGEON REPORTED COULD NOT BE REPEATED. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE IT FUNCTIONS PROPERLY. CO STIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE CO'S PRODUCTS.

Description of Event or Problem · 1

DURING A LAPAROSCOPIC ASSISTED VAGINAL HYSTERECTOMY, THE SURGEON FIRED THE TSW35 SIX TIMES WITHOUT DIFFICULTY. ON THE SEVENTH FIRING THE DEVICE MISFIRED. A SECOND DEVICE WAS OPENED TO COMPLETE THE CASE. THERE WAS NO CONSEQUENCE TO THE PT. CLINICAL FOLLOW-UP: 1/24/97 1340 SURGEON DESCRIBED SEVERAL DIFFICULTIES WITH THE INSTRUMENT: SOMETIMES THE INSTRUMENT "UNLOCKED BEFORE FIRING," SOMETIMES IT WOULD NOT OPEN AFTER FIRING, AND IT FIRED STAPLES BUT DID NOT CUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS LINEAR CUTTER KOG ETHICON ENDO-SURGERY, INC. NA J4500F

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other