FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE
MDR report key: 822223
·
Received January 25, 2007
Report
- Report Number
- 2954323-2007-00391
- Event Type
- Malfunction
- Date Received
- January 25, 2007
- Date of Event
- October 14, 2006
- Report Date
- January 25, 2007
- Manufacturer
- ABBOTT DIABETES CARE INC, USA
- Product Code
- LFR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
COMPLAINT IS CONFIRMED. PERFORMED INVESTIGATION. MEMORY OVERWRITE WAS OBSERVED. FOUND OUM TO BE SELECTABLE AND SET TO MG/DL. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH THE FA16MAY2006 LETTER.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING AN ERROR MESSAGE ON HER LOCKED FREESTYLE METER. THE DEVICE WAS RETURNED AND DURING THE COURSE OF THE INVESTIGATION, IT WAS DISCOVERED THE METER HAD SUFFERED THE MEMORY OVERWRITE MALFUNCTION. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC, USA | NI | 0607353 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |