FDA Adverse Event Other Summary report: N

MTS BUFFERED GEL CARD

MDR report key: 821680 · Received February 12, 2007

Report

Report Number
1056600-2007-00042
Event Type
Other
Date Received
February 12, 2007
Date of Event
January 15, 2007
Report Date
February 12, 2007
Manufacturer
MICRO TYPING SYSTEMS, INC.
Product Code
KSZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MTS BUFFERED GEL CARDS ARE NOT INTENDED FOR USE WITH INDIRECT ANTIGLOBULIN TESTING. THE USER SELF-IDENTIFIED THE ISSUE, CONTACTED OCD AND ALL TESTING WAS REPEATED. A DEFINITIVE ROOT CAUSE WAS DETERMINED. INCIDENT OCCURRED AS A RESULT OF USER ERROR. PRODUCT LABELING CLEARLY INDICATES WHICH TESTS ARE APPROPRIATE FOR USE WITH MTS BUFFERED GEL CARDS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY INADVERTENTLY PERFORMED INDIRECT ANTIGLOBULIN TESTING ON 40 PATIENT SAMPLES WITH MTS BUFFERED GEL CARDS INSTEAD OF ANTI-IGG CARDS. UPON INSTRUCTION FROM OCD; THE CUSTOMER RE-TESTED ALL OF THE PATIENT SAMPLES WITH MTS ANT-IGG GEL CARDS. THE CUSTOMER REPORTED THAT ONE OF THE 40 PATIENT SAMPLES REACTED POSITIVE FOR ANTIBODY UPON RE-TEST. THE CUSTOMER STATED NO TRANSFUSION REACTIONS WERE REPORTED AND AT THIS POINT, THEY WERE UNSURE IF ANY PATIENT HAD RECEIVED A TRANSFUSION. PRODUCT LABELING CLEARLY INDICATES THAT MTS BUFFERED GEL CARDS ARE INTENDED FOR USE WITH THE A/B/O SERUM GROUPING AS WELL AS DIRECT AGGLUTINATION I.E. COLD AND WARM ANTIBODY DETECTION WITH ENZYME TREATED CELLS. MTS BUFFERED GEL CARDS ARE NOT INTENDED FOR USE WITH INDIRECT ANTIGLOBULIN TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MTS BUFFERED GEL CARD ABO GROUP - RBC KSZ MICRO TYPING SYSTEMS, INC. NA 042406004-02

Patients

Seq Age Sex Outcome Treatment
1 *