FDA Adverse Event Malfunction Summary report: N

BD CONNECTA¿ STOPCOCK

MDR report key: 8214584 · Received January 3, 2019

Report

Report Number
9610847-2018-00459
Event Type
Malfunction
Date Received
January 3, 2019
Date of Event
December 17, 2018
Report Date
February 8, 2019
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMG
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: IN RESPONSE TO THE EVENT REPORTED BY YOUR FACILITY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 8151681. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, OUR ENGINEERS WERE ABLE TO DUPLICATE THIS EVENT THROUGH THE LEAKAGE TESTING OF THE SUBMITTED DEVICE. A SUBSEQUENT REVIEW OF OUR MANUFACTURING LINE DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR THIS EVENT IS AN ABNORMALITY IN THE EQUIPMENT RESPONSIBLE FOR TUBING ASSEMBLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD CONNECTA¿ STOPCOCK INJECTION VALVE LEAKS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

(B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD CONNECTA¿ STOPCOCK INJECTION VALVE LEAKS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8032 BD CONNECTA¿ STOPCOCK STOPCOCK FMG BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 8151681

Patients

Seq Age Sex Outcome Treatment
1 Other