BD PHASEAL¿ PROTECTOR P50J
Report
- Report Number
- 3003152976-2018-00569
- Event Type
- Malfunction
- Date Received
- December 31, 2018
- Date of Event
- December 11, 2018
- Report Date
- January 22, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- LHI
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY: ONE SAMPLE UNIT WAS RECEIVED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH VISUAL INSPECTION OF THE SAMPLE, A PIECE OF HAIR WAS FOUND WITHIN THE EXPANSION CHAMBER. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR THE PROVIDED LOT NUMBER (1804125) AND THE REVIEW DID NOT REVEAL ANY QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS INCIDENT. PHASEAL PRODUCTS ARE MANUFACTURING AND ASSEMBLED IN A DESIGNATED CLEAN AREA. PERSONNEL ARE REQUIRED TO FOLLOW STRICT GOWNING AND PREVENTIVE PROCEDURES PRIOR TO ENTERING THE CLEAN AREA. TO IMPROVE THE CLEANING OF THE LINES, A PROJECT WAS OPEN ON FY¿18 (#1436). THE PROJECT INCLUDED NEW CLEANING FREQUENCIES AND RE-TRAINING OF THE PERSONNEL IN THE AREA. THE PROJECT HAS DECREASED THE NUMBER OF COMPLAINTS RELATED TO FM. THE LOT WAS MANUFACTURED WHEN THE IMPROVEMENTS WERE BEING LAUNCHED. PERSONNEL OF THE AREA HAS BEEN AWARE OF THE COMPLAINT.
IT WAS REPORTED THAT "HAIR LIKE" FOREIGN MATTER WAS FOUND IN THE BD PHASEAL¿ PROTECTOR P50J.
(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT "HAIR LIKE" FOREIGN MATTER WAS FOUND IN THE BD PHASEAL¿ PROTECTOR P50J.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1045950 | BD PHASEAL¿ PROTECTOR P50J | PHASEAL ADMINISTRATION SET | LHI | BECTON DICKINSON, S.A. | 1804125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |