FDA Adverse Event
Other
Summary report: N
I-STAT CTNI CARTRIDGE
MDR report key: 820419
·
Received February 9, 2007
Report
- Report Number
- 2245578-2007-00005
- Event Type
- Other
- Date Received
- February 9, 2007
- Date of Event
- January 22, 2007
- Report Date
- February 9, 2007
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
IN 2007, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT I-STAT CARDIAC TROPONIN I CARTRIDGE YIELDED A RESULT OF 0.32 NG/ML. A SAMPLE WHICH WAS DRAWN AT THE SAME TIME, YIELDED A CARDIAC TROPONIN I RESULT OF 0.01 NG/ML ON A LAB ANALYZER. THE I-STAT RESULT OF 0.32 NG/ML CONTRIBUTED TO THE DECISION TO SENT THE PT FOR CARDIAC CATHETERIZATION. TEST RESULTS FROM THE CARDIAC CATHETERIZATION WERE NEGATIVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT CTNI CARTRIDGE | IVD | MMI | ABBOTT POINT OF CARE INC. | NA | U06283A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR | Other |