TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 6000093-2007-00241
- Event Type
- Malfunction
- Date Received
- February 1, 2007
- Date of Event
- December 26, 2006
- Report Date
- January 23, 2007
- Manufacturer
- BOSTON SCIENTIFIC CORP.
- Product Code
- NIQ
- PMA / PMN Number
- P030025
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVAL: PRODUCT ANALYSIS COULD NEITHER CONFIRM NOR DENY THE CROSSING DIFFICULTY AS STATED IN THE COMPLAINT. VISUAL EXAMINATION OF THE STENT REVEALED DAMAGE TO THE PROXIMAL END. THE PROXIMAL STENT STRUTS WERE BENT. THE BALLOON WAS VISUALLY AND MICROSCOPICALLY EXAMINED. NO VISUAL DEFECTS WERE OBSERVED UNDER MAGNIFICATION. THE BALLOON WAS TIGHTLY WRAPPED AND DID NOT APPEAR TO HAVE BEEN SUBJECTED TO ANY POSITIVE PRESSURE. THERE IS NO EVIDENCE THAT THE DEVICE WAS IMPROPERLY MANUFACTURED. DAMAGE OF THIS NATURE IS USUALLY THE RESULT OF PULLING AN UN-EXPANDED STENT BACK INTO THE GUIDE CATHETER. THE EXACT CAUSE OF THE STENT DAMAGE COULD NOT BE DETERMINED. THE MFG RECORDS FOR BATCH 8232905 HAVE BEEN REVIEWED AND NO ISSUES OR DISCREPANCIES WERE FOUND. THIS RECORDS REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PERFORMANCE SPECIFICATIONS AT THE TIME OF ITS RELEASE TO DISTRIBUTION.
THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON 01/23/2007. IT WAS REPORTED THAT DURING A CORONARY STENTING TREATMENT PROCEDURE, THE STENT COULD NOT CROSS THE LESION. AFTER PRE-DILATING THE SEVERELY CALCIFIED AND MODERATELY TORTUOUS RIGHT CORONARY ARTERY, THE PHYSICIAN ATTEMPTED TO PLACE A 4.50 X 16MM TAXUS EXPRESS2 DRUG ELUTING STENT, BUT HAD DIFFICULTY CROSSING THE LESION. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THE PT'S CONDITION IS REPORTED AS "GOOD". HOWEVER, THE RETURNED PRODUCT CONFIRMED THAT THERE WAS PROXIMAL STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM | NIQ STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC CORP. | 4.50 X 16MM | 8232905 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |