HEMOSPRAY ENDOSCOPIC HEMOSTAT
Report
- Report Number
- 1037905-2018-00598
- Event Type
- Malfunction
- Date Received
- December 21, 2018
- Date of Event
- November 22, 2018
- Report Date
- November 26, 2018
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- QAU
- UDI-DI
- 00827002213463
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
INFORMATION REGARDING SUSPECT MEDICAL DEVICE; COMMON DEVICE NAME: NOT AVAILABLE; REGULATION NAME: HEMOSTATIC DEVICE FOR INTRALUMINAL GASTROINTESTINAL USE. DEN170015. INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED CONFIRMED THE REPORT. HOWEVER, ALL THE COMPONENTS WERE NOT INCLUDED IN THE RETURN (NO RETURNED CATHETERS) THEREFORE A COMPLETE EVALUATION WAS NOT POSSIBLE. THE DEVICE WAS RETURNED WITH THE ON/OFF SWITCH IN THE "OFF" POSITION WITH THE RED ACTIVATION KNOB ENGAGED IN THE HANDLE FOR ACTIVATION OF THE CARBON DIOXIDE CARTRIDGE. THE POWDER CHAMBER APPEARED TO BE RATHER FULL. WHEN TESTED AS RETURNED, THE DEVICE DID NOT SPRAY. THE RED ACTIVATION HANDLE WAS REMOVED, NO CARBON DIOXIDE HAD REMAINED IN THE CARTRIDGE. THE CARBON DIOXIDE CARTRIDGE WAS PUNCTURED APPROPRIATELY. THE LANCE POSITION WAS MEASURED USING A TOOL AND FOUND TO BE POSITIONED CORRECTLY. A VISUAL EXAMINATION OF THE LANCE INSIDE THE HANDLE INDICATES IT IS BEVELED. THE O-RING AND LANCE APPEAR TO BE POSITIONED CORRECTLY INSIDE THE REGULATOR. WHEN RETESTED WITH A NEW CARBON DIOXIDE CARTRIDGE, A PUFF OF POWDER WAS EXPELLED WHEN THE SWITCHED WAS TOGGLED TO THE "ON" POSITION. THE DEVICE THEN SPRAYED POWDER AS INTENDED. A PRODUCT DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORTED OCCURRENCE WAS NOT OBSERVED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSION: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. FURTHERMORE, WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE CATHETERS WERE NOT RETURNED FOR EVALUATION. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. THE INSTRUCTIONS FOR USE STATES: "IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITIONS, DO NOT USE." THE REPORT INDICATES, "IT WAS NOT POSSIBLE TO SPRAY ALSO BEFORE THE DEVICE HAS BEEN PLACED IN THE ENDOSCOPE." THE EVALUATION AND REPORT INDICATE THE DEVICE WAS USED. THE REPORT SUGGESTS THE USER TESTED THE DEVICE PRIOR TO INSERTING THE CATHETER INTO THE ENDOSCOPE, DETERMINED THE DEVICE DID NOT SPRAY OUTSIDE OF THE ENDOSCOPE, AND THEN ATTEMPTED TO USE THE NONFUNCTIONING DEVICE IN THE ENDOSCOPE. CATHETER OCCLUSION CAN BE A CAUSE OF THIS DEVICE NOT BEING ABLE TO SPRAY POWDER. CATHETER OCCLUSION CAN BE RELATED TO THE HANDLING OF THE DEVICE DURING THE PROCEDURE. TO ASSIST THE USER, THE INSTRUCTIONS FOR USE STATES THE FOLLOWING: "NOTE: DO NOT TEST DEVICE PRIOR TO INSERTION INTO ENDOSCOPE ACCESSORY CHANNEL AS THIS MAY INCREASE RISK OF CATHETER OCCLUSION... PRECAUTION: TO AVOID CATHETER OCCLUSION, DO NOT PLACE CATHETER DIRECTLY IN CONTACT WITH BLOOD AND/OR MUCOSA, INCLUDING ANY POOLED BLOOD AND DO NOT ASPIRATE BLOOD WHILE CATHETER IS IN ACCESSORY CHANNEL... NOTE: IF CATHETER BECOMES OCCLUDED, TURN RED VALVE TO CLOSED POSITION, REMOVE CATHETER FROM ENDOSCOPE AND REPLACE WITH EXTRA CATHETER PROVIDED IN PACKAGE." PRIOR TO DISTRIBUTION, ALL HEMOSPRAY ENDOSCOPIC HEMOSTATS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.
DURING A HEMOSTASIS PROCEDURE, THE PHYSICIAN USED A COOK HEMOSPRAY ENDOSCOPIC HEMOSTAT. THE CARTRIDGE COULD NOT BE ACTIVATED CORRECTLY. THEY ACTIVATED THE CARBON DIOXIDE CARTRIDGE BUT COULD NOT SPRAY [POWDER]. PER THE COOK SALES REPRESENTATIVE, "I SUSPECT THAT THE CARBON DIOXIDE CARTRIDGE HAS NOT OPENED [BEEN PUNCTURED]. IT WAS NOT POSSIBLE TO SPRAY, ALSO BEFORE THE DEVICE HAS BEEN PLACED IN THE ENDOSCOPE [DEVICE ALSO DID NOT SPRAY PRIOR TO USE]." A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1031759 | HEMOSPRAY ENDOSCOPIC HEMOSTAT | QAU | COOK ENDOSCOPY | W4017240 | 00827002213463 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |