FDA Adverse Event Malfunction Summary report: N

SERVO-I

MDR report key: 8189093 · Received December 21, 2018

Report

Report Number
8010042-2018-00697
Event Type
Malfunction
Date Received
December 21, 2018
Report Date
January 24, 2019
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K123149
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4). THE VENTILATOR WAS INVESTIGATED ON SITE AND THE SOFTWARE WAS RELOADED. THE DC/DC & STANDARD CONNECTORS PRINTED CIRCUIT BOARD (PCB) WAS REPLACED AND RETURNED FOR INVESTIGATION. THE DC/DC & STANDARD CONNECTORS PCB WAS INSTALLED IN A VENTILATOR FOR SIMULATED USE TESTING. NO ALARMS WERE GENERATED OR OTHER PROBLEM OCCURRED DURING THE TEST. THE DC/DC & STANDARD CONNECTORS PCB WAS WORKING AS EXPECTED. EVALUATION OF THE RECEIVED LOG FILE CONFIRMS THE BOOT UP PROBLEM. BASED ON THE DEVICE LOG EVALUATION OUR CONCLUSION IS THAT THE PROBLEM WAS LIKELY CAUSED BY AN INTERNAL COMMUNICATION ERROR AND IT WAS RESOLVED BY A RE-INSTALLATION OF THE SAME SOFTWARE VERSION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VENTILATOR DID NOT START. THERE WAS NO PATIENT INVOLVEMENT. (B)(4).

Description of Event or Problem · 0

IMPORTER REF. #: CC-CPL-2018-01928. MANUFACTURER REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1028525 SERVO-I VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1