UNKNOWN CATHETER
Report
- Report Number
- 2021898-2018-00580
- Event Type
- Injury
- Date Received
- December 21, 2018
- Report Date
- December 20, 2018
- Manufacturer
- MEDTRONIC NEUROSURGERY
- Product Code
- JXG
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
MICHAEL KIM, ILYA RYBKIN, HARRISON SMITH, JARED COOPER, MICHAEL TOBIAS. BONE OVERGROWTH CAUSING PROXIMAL VENTRICULOPERITONEAL SHUNT MALFUNCTION. WORLD NEUROSURGERY 121 (2019). DOI: 10.1016/J.WNEU.2018.10.030. BACKGROUND: HYDROCEPHALUS IS AN INTERNATIONAL DISEASE PROCESS THAT IS COMMONLY TREATED SURGICALLY WITH A VENTRICULOPERITONEAL SHUNT. THIS DEVICE MAY BE PRONE TO MALFUNCTION, MOST COMMONLY FROM OBSTRUCTION, DISCONNECTION, OR INFECTION. CASE DESCRIPTION: A (B)(6) YEAR-OLD FEMALE WITH HYDROCEPHALUS AND A VENTRICULOPERITONEAL SHUNT PRESENTED WITH ALTERED MENTAL STATUS AND IMAGING CONCERNING FOR A SHUNT MALFUNCTION. INTRAOPERATIVELY, SHE WAS FOUND TO HAVE BONE GROWING OVER AND COMPRESSING THE PROXIMAL OCCLUDER OF THE SHUNT VALVE, CAUSING A MECHANICAL OBSTRUCTION. REMOVAL OF THE BONE ALLOWED FOR EGRESS OF CEREBROSPINAL FLUID AND RETURN OF PROPER SHUNT FUNCTION. THE PATIENT DID WELL POSTOPERATIVELY. CONCLUSION: HYDROCEPHALUS, VENTRICULOPERITONEAL SHUNTS, AND SHUNT REVISIONS REPRESENT A SIGNIFICANT HEALTH BURDEN AND COST. HERE WE PRESENT AN UNUSUAL CAUSE OF A SHUNT MALFUNCTION CAUSED BY BONY OVERGROWTH. REPORTED EVENTS. THE (B)(6) YEAR OLD FEMALE HAD ONE PRIOR REVISION OF THEIR VPS 8 YEARS PRIOR DUE TO FRACTURING OF THE DISTAL CATHETER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1028072 | UNKNOWN CATHETER | SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS | JXG | MEDTRONIC NEUROSURGERY | UNKNOWN-C | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Hospitalization| R |