FDA Adverse Event
Injury
Summary report: N
GENTAL POWER LUX CONTRA ANGLE
MDR report key: 818746
·
Received February 15, 2007
Report
- Report Number
- 1419798-2007-00004
- Event Type
- Injury
- Date Received
- February 15, 2007
- Date of Event
- January 26, 2007
- Report Date
- January 29, 2007
- Manufacturer
- KAVO AMERICA
- Product Code
- EKK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE DOCTOR WAS INFORMED THAT THE HANDPIECE WAS DENTED AND BEARINGS AND GEARS ARE WORN AND GRITTY. IN ADDITION, MAINTENANCE INFO WAS REVIEWED WAS WITH DOCTOR'S OFFICE.
Description of Event or Problem · 1
THE HEAD OF HANDPIECE BECAME HOT AND BURNED THE PT. THE BURN WAS THE SIZE OF BACK CAP. OINTMENT WAS PRESCRIBED. SHE HEALED FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTAL POWER LUX CONTRA ANGLE | * | EKK | KAVO AMERICA | 25LPA | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Required Intervention |