FDA Adverse Event Injury Summary report: N

GENTAL POWER LUX CONTRA ANGLE

MDR report key: 818741 · Received February 15, 2007

Report

Report Number
1419798-2007-00003
Event Type
Injury
Date Received
February 15, 2007
Date of Event
November 15, 2006
Report Date
January 11, 2007
Manufacturer
KAVO AMERICA
Product Code
EKK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE DOCTOR WAS INFORMED THAT THE HANDPIECE BEARINGS ARE WORN AND GRITTY. A SERVICE CHECK WAS RECOMMENDED TO EVALUATE IF IT WAS TIME FOR MAINTENANCE REPAIR. IN ADDITION, MAINTENANCE INFO WAS REVIEWED WITH THE DR.

Description of Event or Problem · 1

THE HEAD OF HANDPIECE BECAME HOT AND BURNED THE PT. THE INITIAL BURN CREATED A SMALL BLISTER. OINTMENT WAS APPLIED TO THE BURN. SHE HEALED FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTAL POWER LUX CONTRA ANGLE * EKK KAVO AMERICA 25LPA *

Patients

Seq Age Sex Outcome Treatment
1 NA Required Intervention