FDA Adverse Event Injury Summary report: N

GENTALPOWER LUX CONTR ANGLE

MDR report key: 818739 · Received February 15, 2007

Report

Report Number
1419798-2007-00002
Event Type
Injury
Date Received
February 15, 2007
Date of Event
January 16, 2007
Report Date
January 18, 2007
Manufacturer
KAVO AMERICA
Product Code
EKK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE DOCTOR WAS INFORMED THAT THERE WAS A DENT IN THE EDGE OF THE HEAD OF THE HANDPIECE THAT CAUSED THE HEAD TO HEAT UP. IN ADDITION, MAINTENANCE INFO AND SETTINGS WERE REVIEWED WITH THE DR. MAINTENANCE SHEETS WERE SENT.

Description of Event or Problem · 1

THE HEAD OF THE HANDPIECE BECAME HOT AND BURNED THE PT. ANTI-INFLAMMATORY GEL WAS APPLIED TO THE BURN. HE HEALED FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTALPOWER LUX CONTR ANGLE * EKK KAVO AMERICA 25LPA *

Patients

Seq Age Sex Outcome Treatment
1 NA Required Intervention