FDA Adverse Event Injury Summary report: N

NSK

MDR report key: 8186552 · Received December 20, 2018

Report

Report Number
9611253-2018-00074
Event Type
Injury
Date Received
December 20, 2018
Date of Event
July 1, 2015
Report Date
August 8, 2019
Manufacturer
NAKANISHI INC.
Product Code
KMW
PMA / PMN Number
K171155
Removal / Correction Number
9611253-060818-001-R
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016004. (B)(4). NAKANISHI IS CURRENTLY TRYING TO OBTAIN FURTHER INFORMATION ABOUT THE EVENT AND PATIENT.

Additional Manufacturer Narrative · 0

NSK AMERICA (NAM) TOOK THE FOLLOWING ACTIONS, BUT FURTHER INFORMATION ABOUT THE EVENT AND PATIENT WAS NOT PROVIDED BY THE DENTIST. - ON DECEMBER 14, 2018, NAM MADE A PHONE CALL TO THE DENTAL OFFICE AND E-MAILED AN INFORMATION FORM (QA-036). - ON DECEMBER 19, 2018, NAM SENT AN E-MAIL TO THE DENTIST. - ON JANUARY 29, 2019, NAM SENT ANOTHER E-MAIL TO THE DENTIST.

Additional Manufacturer Narrative · 0

ON JULY 20, 2019, NAKANISHI WAS INFORMED BY NAM THAT THE FOLLOWING ACTIONS HAD BEEN TAKEN.. IN NOVEMBER 2018 (EXACT DATE UNKNOWN), A DISTRIBUTOR SENT THE UPDATED OPERATION MANUALS (PROVIDED TO THEM BY NAM) THAT CONTAIN THE DETAILED REPROCESSING METHOD TO THE DENTAL OFFICE TO PREVENT THE DEVICES FROM OVERHEATING. ON DECEMBER 5, 2018, NAM CONFIRMED THE RECEIPT OF THE OPERATION MANUALS BY THE DENTIST VIA E-MAIL FROM THE DISTRIBUTOR.

Description of Event or Problem · 0

ON (B)(6) 2018, NAKANISHI RECEIVED AN E-MAIL FROM A DISTRIBUTOR (B)(4) ABOUT A HANDPIECE OVERHEATING. ACCORDING TO (B)(4), THE INFORMATION WAS PROVIDED THROUGH THE ADVERSE EVENT QUESTIONNAIRE ABOUT THE SGS/SGA HANDPIECES AS A PART OF RECALL ACTIVITIES FOR THOSE DEVICES. THE DETAILS ARE AS FOLLOWS. THE EVENT OCCURRED IN (B)(6) 2015. (THE EXACT DATE IS UNKNOWN). A DENTIST WAS PERFORMING A DENTAL PROCEDURE USING THE SGS-E2S HANDPIECE (SERIAL NO. UNKNOWN). DURING THE PROCEDURE, THE HANDPIECE OVERHEATED AND BURNED A PATIENT'S LIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1026499 NSK HANDPIECE, ROTARY BONE CUTTING KMW NAKANISHI INC. SGS-E2S

Patients

Seq Age Sex Outcome Treatment
1 Other