FDA Adverse Event Malfunction Summary report: N

GRYPHON P BR DS ANCHOR W/OC

MDR report key: 8182370 · Received December 19, 2018

Report

Report Number
1221934-2018-55786
Event Type
Malfunction
Date Received
December 19, 2018
Date of Event
December 8, 2018
Report Date
March 13, 2019
Manufacturer
DEPUY MITEK LLC US
Product Code
MAI
UDI-DI
10886705001279
PMA / PMN Number
K150209
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RP
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY : THE COMPLAINT DEVICE IS NOT BEING RETURNED, IT WAS DISCARDED BY THE CUSTOMER, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. HENCE, THIS COMPLAINT CANNOT BE CONFIRMED. NO FURTHER INFORMATION REGARDING THE TECHNIQUE OR INSTRUMENTS USED HAS BEEN PROVIDED TO DETERMINE A ROOT CAUSE FOR THIS FAILURE. IF ANY ADDITIONAL INFORMATION IS OBTAINED, THIS COMPLAINT WILL BE RE-OPENED TO CAPTURE THAT INFORMATION. FURTHERMORE, A NON-CONFORMANCE SEARCH WAS CONDUCTED AND NO NON-CONFORMANCES WERE IDENTIFIED FOR THIS PART NUMBER (210813), LOT NUMBER (L189929) COMBINATION. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. UDI: (B)(4).

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE COMPLAINT DEVICE WAS RECEIVED AND INSPECTED. THE COMPLAINT ALLEGING THE MIGRATION OF THE DEVICE OUT OF THE BONE HOLE CAN BE CONFIRMED. IT WAS OBSERVED THAT THE DEVICE HAD TISSUE DEBRIS COVERING THE THREADS, AND A PIECE OF THE PROXIMAL END OF THE DEVICE HAD BEEN BROKEN OFF. ANCHOR BREAKAGES ARE TYPICALLY ASSOCIATED WITH OFF AXIS INSERTION, LEVERING DURING INSERTION, HARD BONE QUALITY OR USING INCORRECT INSTRUMENTATION FOR PREPARING BONE HOLE. THE DEVICE MAY HAVE MIGRATED OUT OF THE BONE HOLE DUE TO THE PROXIMAL END OF THE DEVICE BEING BROKEN THEREFORE CAUSING INSTABILITY IN THE SEATING OF THE DEVICE. HOWEVER, GIVEN THE INFORMATION PROVIDED WE CANNOT DISCERN A DEFINITIVE ROOT CAUSE FOR THE REPORTED FAILURE. A NON-CONFORMANCE SEARCH WAS CONDUCTED TO INVESTIGATE ANY DEFECTS IDENTIFIED DURING PRODUCTION THAT MAY CONTRIBUTE TO THE COMPLAINT CONDITION. NO NON-CONFORMANCE WAS IDENTIFIED FOR THIS PART-LOT NUMBER COMBINATION. AT THIS POINT, NO CORRECTIVE ACTION IS REQUIRED AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. CORRECTED DATA (H1): UPON FURTHER REVIEW OF THE COMPLAINT, IT WAS DETERMINED THAT THE PREVIOUSLY REPORTED SERIOUS INJURY IS UNLIKELY TO CAUSE OR CONTRIBUTE TO PATIENT HARM. THEREFORE, THIS COMPLAINT IS DETERMINED TO BE A MALFUNCITON. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY. UDI: (B)(4).

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT #: (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY: THE COMPLAINT DEVICE IS NOT BEING RETURNED, IT WAS DISCARDED BY THE CUSTOMER, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. HENCE, THIS COMPLAINT CANNOT BE CONFIRMED. NO FURTHER INFORMATION REGARDING THE TECHNIQUE OR INSTRUMENTS USED HAS BEEN PROVIDED TO DETERMINE A ROOT CAUSE FOR THIS FAILURE. IF ANY ADDITIONAL INFORMATION IS OBTAINED, THIS COMPLAINT WILL BE RE-OPENED TO CAPTURE THAT INFORMATION. FURTHERMORE, A NON-CONFORMANCE SEARCH WAS CONDUCTED AND NO NON-CONFORMANCES WERE IDENTIFIED FOR THIS PART NUMBER (210813), LOT NUMBER (L189929) COMBINATION. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE VIA EMAIL ON 01/11/2019 STATES THAT AN ADVERSE EVENT OCCURRED DURING THE PROCEDURE ON (B)(6) 2018. THE AFFILIATE DESCRIBES THE ADVERSE EVENT AS THE ANCHOR PULLING OUT FROM THE BONE. THE PROCEDURE WAS COMPLETED BY USING A HEALIX 4.5MM. UDI: (B)(4).

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY: THE COMPLAINT DEVICE HAS NOT BEEN RETURNED, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. HENCE, THIS COMPLAINT CANNOT BE CONFIRMED. NO FURTHER INFORMATION REGARDING THE TECHNIQUE OR INSTRUMENTS USED HAS BEEN PROVIDED TO DETERMINE A ROOT CAUSE FOR THIS FAILURE. IF ANY ADDITIONAL INFORMATION IS OBTAINED, THIS COMPLAINT WILL BE RE-OPENED TO CAPTURE THAT INFORMATION. FURTHERMORE, A NON-CONFORMANCE SEARCH WAS CONDUCTED AND NO NON-CONFORMANCES WERE IDENTIFIED FOR THIS PART NUMBER (210813), LOT NUMBER (L189929) COMBINATION. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. UDI: (B)(4).

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY : THE COMPLAINT DEVICE IS NOT BEING RETURNED, IT WAS DISCARDED BY THE CUSTOMER, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. HENCE, THIS COMPLAINT CANNOT BE CONFIRMED. NO FURTHER INFORMATION REGARDING THE TECHNIQUE OR INSTRUMENTS USED HAS BEEN PROVIDED TO DETERMINE A ROOT CAUSE FOR THIS FAILURE. IF ANY ADDITIONAL INFORMATION IS OBTAINED, THIS COMPLAINT WILL BE RE-OPENED TO CAPTURE THAT INFORMATION. FURTHERMORE, A NON-CONFORMANCE SEARCH WAS CONDUCTED AND NO NON-CONFORMANCES WERE IDENTIFIED FOR THIS PART NUMBER (210813), LOT NUMBER (L189929) COMBINATION. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. UDI: (B)(4).

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE VIA EMAIL ON 01/11/2019 STATES THAT AN ADVERSE EVENT OCCURRED DURING THE PROCEDURE ON (B)(6) 2018. THE AFFILIATE DESCRIBES THE ADVERSE EVENT AS THE ANCHOR PULLING OUT FROM THE BONE. THE PROCEDURE WAS COMPLETED BY USING A HEALIX 4.5MM.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED FROM AFFILIATE STATING THAT THE EVENT OCCURRED INTRA-OPERATIVE AND THERE WAS A DELAY OF 5 MINUTES. THERE WAS NO PATIENT IMPACT DUE TO THE DELAY.THE SAME BONE HOLE WAS USED TO COMPLETE THE PROCEDURE. A DIFFERENT READILY AVAILABLE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE ADDITIONAL INFORMATION ALSO STATED THAT THE DEVICE IS AVAILABLE FOR RETURN.

Additional Manufacturer Narrative · 1

IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY. UDI: (B)(4). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED BY THE AFFILIATE IN THE (B)(6) THAT DURING A BANKART REPAIR SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE TIP ON THE GRYPHON P BR DS ANCHOR W/OC DEVICE WAS DEFORMED. IT WAS NOT REPORTED IF THERE WAS A DELAY IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1018150 GRYPHON P BR DS ANCHOR W/OC SOFT-TISSUE ANCHOR, BIOABSORBABLE MAI DEPUY MITEK LLC US L189929 10886705001279

Patients

Seq Age Sex Outcome Treatment
1