Narratives
INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR DOUBLE LABEL WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION/TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND DOUBLE LABEL WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE CUSTOMER PHOTOS, THE CUSTOMER¿S INDICATED FAILURE MODE FOR DOUBLE LABEL WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION OF THE CUSTOMER SAMPLES WAS CONDUCTED DOUBLE LABEL OBSERVED. ROOT CAUSE DESCRIPTION: BASED ON THE INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED. RATIONALE: COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT A TUBE URIN PLH 16X100 BC RD HAD TWO DIFFERENT BARCODE STICKERS ON IT.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT A TUBE URIN PLH 16X100 BC RD HAD TWO DIFFERENT BARCODE STICKERS ON IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1017021 | TUBE URIN PLH 16X100 BC RD | COLLECTION DEVICE | JXA | BECTON, DICKINSON AND COMPANY (BD) | 7325913 |