FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 8179193 · Received December 19, 2018

Report

Report Number
1024879-2018-01493
Event Type
Malfunction
Date Received
December 19, 2018
Date of Event
December 3, 2018
Report Date
April 1, 2019
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: HEMOLYSIS CAN BE CAUSED BY MANY SOURCES, INCLUDING CERTAIN PATIENT PATHOLOGICAL CONDITIONS, IMPROPER SPECIMEN COLLECTION, SPECIMEN PROCESSING, AND SPECIMEN TRANSPORT. SPECIMEN COLLECTION FACTORS THAT CAN CONTRIBUTE TO HEMOLYSIS RANGE FROM PROLONGED TOURNIQUET TIME AND IMPROPER VENIPUNCTURE TECHNIQUE TO TRANSFERRING A SAMPLE FROM A SYRINGE DRAW OR IV CATHETER. SPECIMEN PROCESSING FACTORS INCLUDE VIGOROUS MIXING OR SHAKING OF THE SPECIMEN, NOT ALLOWING THE SPECIMEN TO CLOT FOR THE RECOMMENDED AMOUNT OF TIME, PROLONGED CONTACT OF SERUM OR PLASMA WITH CELLS, AND EXPOSURE TO EXCESSIVE HEAT OR COLD. FINALLY, MECHANICAL TRAUMA AND OTHER ADVERSE CONDITIONS DURING TRANSPORT OF TUBES CAN RESULT IN HEMOLYSIS. ERRONEOUS POTASSIUM RESULTS CAN BE CAUSED BY MANY SOURCES, INCLUDING CERTAIN PATIENT PATHOLOGICAL CONDITIONS, IMPROPER SPECIMEN COLLECTION, SPECIMEN PROCESSING, AND SPECIMEN TRANSPORT. SPECIMEN COLLECTION FACTORS THAT CAN CONTRIBUTE TO ERRONEOUS POTASSIUM RANGE FROM PROLONGED TOURNIQUET TIME AND IMPROPER VENIPUNCTURE TECHNIQUE TO TRANSFERRING A SAMPLE FROM A SYRINGE DRAW OR IV CATHETER. SPECIMEN PROCESSING FACTORS INCLUDE VIGOROUS MIXING OR SHAKING OF THE SPECIMEN, NOT ALLOWING THE SPECIMEN TO CLOT FOR THE RECOMMENDED AMOUNT OF TIME, PROLONGED CONTACT OF SERUM OR PLASMA WITH CELLS, AND EXPOSURE TO EXCESSIVE HEAT OR COLD. FINALLY, MECHANICAL TRAUMA AND OTHER ADVERSE CONDITIONS DURING TRANSPORT OF TUBES CAN RESULT IN ERRONEOUS POTASSIUM RESULTS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH 8166531, REORDER NUMBER (B)(4). PRODUCT WAS MANUFACTURED AT THE BD SUMTER SITE JUNE, 2018. PRODUCT QUALITY IS EVALUATED DURING THE MANUFACTURING PROCESS WITH PRESCRIBED VARIABLE AND ATTRIBUTES INSPECTIONS. ALL INSPECTIONS WERE IN COMPLIANCE WITH REQUIREMENTS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH 8120676, REORDER NUMBER (B)(4). PRODUCT WAS MANUFACTURED AT THE BD SUMTER SITE MAY, 2018. PRODUCT QUALITY IS EVALUATED DURING THE MANUFACTURING PROCESS WITH PRESCRIBED VARIABLE AND ATTRIBUTES INSPECTIONS. ALL INSPECTIONS WERE IN COMPLIANCE WITH REQUIREMENTS.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. REGARDING ERRONEOUS RESULTS / HEMOLYSIS: HEMOLYSIS CAN BE CAUSED BY MANY SOURCES, INCLUDING CERTAIN PATIENT PATHOLOGICAL CONDITIONS, IMPROPER SPECIMEN COLLECTION, SPECIMEN PROCESSING, AND SPECIMEN TRANSPORT. SPECIMEN COLLECTION FACTORS THAT CAN CONTRIBUTE TO HEMOLYSIS RANGE FROM PROLONGED TOURNIQUET TIME AND IMPROPER VENIPUNCTURE TECHNIQUE TO TRANSFERRING A SAMPLE FROM A SYRINGE DRAW OR IV CATHETER. SPECIMEN PROCESSING FACTORS INCLUDE VIGOROUS MIXING OR SHAKING OF THE SPECIMEN, NOT ALLOWING THE SPECIMEN TO CLOT FOR THE RECOMMENDED AMOUNT OF TIME, PROLONGED CONTACT OF SERUM OR PLASMA WITH CELLS, AND EXPOSURE TO EXCESSIVE HEAT OR COLD. FINALLY, MECHANICAL TRAUMA AND OTHER ADVERSE CONDITIONS DURING TRANSPORT OF TUBES CAN RESULT IN HEMOLYSIS. ERRONEOUS POTASSIUM RESULTS CAN BE CAUSED BY MANY SOURCES, INCLUDING CERTAIN PATIENT PATHOLOGICAL CONDITIONS, IMPROPER SPECIMEN COLLECTION, SPECIMEN PROCESSING, AND SPECIMEN TRANSPORT. SPECIMEN COLLECTION FACTORS THAT CAN CONTRIBUTE TO ERRONEOUS POTASSIUM RANGE FROM PROLONGED TOURNIQUET TIME AND IMPROPER VENIPUNCTURE TECHNIQUE TO TRANSFERRING A SAMPLE FROM A SYRINGE DRAW OR IV CATHETER. SPECIMEN PROCESSING FACTORS INCLUDE VIGOROUS MIXING OR SHAKING OF THE SPECIMEN, NOT ALLOWING THE SPECIMEN TO CLOT FOR THE RECOMMENDED AMOUNT OF TIME, PROLONGED CONTACT OF SERUM OR PLASMA WITH CELLS, AND EXPOSURE TO EXCESSIVE HEAT OR COLD. FINALLY, MECHANICAL TRAUMA AND OTHER ADVERSE CONDITIONS DURING TRANSPORT OF TUBES CAN RESULT IN ERRONEOUS POTASSIUM RESULTS. AS NO SAMPLES OR PHOTOS WERE RECEIVED FOR EVALUATION, THE CUSTOMER'S INDICATED FAILURE MODE WAS NOT OBSERVED BY BD. AS THERE WAS NO SAMPLE OR PHOTO AVAILABLE FOR EVALUATION, A ROOT CAUSE COULD NOT BE DETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE THE TUBES WERE POPPING OFF, NOT FILLING, HEMOLYZED, AND HAVING ERRONEOUS RESULTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE THE TUBES WERE POPPING OFF, NOT FILLING, HEMOLYZED, AND HAVING ERRONEOUS RESULTS.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8166531. MEDICAL DEVICE EXPIRATION DATE: 2019-06-30. DEVICE MANUFACTURE DATE: 2018-06-15. MEDICAL DEVICE LOT #: 8120676. MEDICAL DEVICE EXPIRATION DATE: 2019-04-30. DEVICE MANUFACTURE DATE: 2018-04-30. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE THE TUBES WERE POPPING OFF, NOT FILLING, HEMOLYZED, AND HAVING ERRONEOUS RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1017154 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) SEE H.10 50382903679862

Patients

Seq Age Sex Outcome Treatment
1 Other