FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY

MDR report key: 8179189 · Received December 19, 2018

Report

Report Number
1213809-2018-00908
Event Type
Malfunction
Date Received
December 19, 2018
Date of Event
December 4, 2018
Report Date
March 14, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096054
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: TWO LOOSE CONVENIENCE TRAYS IN SEPARATE PLASTIC ZIP LOCK BAGS WERE RETURNED AND VISUALLY EVALUATED. ONE TRAY, REPORTED TO BE FROM BATCH #8211649, WAS EMPTY OF SYRINGES AND OBSERVED TO HAVE A LOOSE BLACK HAIR ON THE BOTTOM OF THE TRAY APPROXIMATELY 1¿ LONG ¿ LARGER THAN LEVEL 3 IN SIZE, WHICH IS REJECTABLE PER PRODUCT SPECIFICATION. ONE TRAY, REPORTED TO BE FROM BATCH #8236598, WAS FILLED WITH 10ML SYRINGES. ONE SYRINGE WAS FOUND TO HAVE DAMAGED FLANGE AND SIDE OF THE BARREL. THE DAMAGE APPEARED TO HAVE SCRAPED PART OF THE FLANGE OFF, WHICH WAS STILL ATTACHED TO THE OPPOSITE SIDE OF THE FLANGE IN BALL OF FUZZY PLASTIC. POTENTIAL ROOT CAUSE FOR THE DAMAGED BARREL DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE TWO PROVIDED LOT NUMBER(S) THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BEFORE USE OF THE BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY A HAIR WAS FOUND INSIDE THE PACKAGING AND THE FLANGE WAS DAMAGED.

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8211649. MEDICAL DEVICE EXPIRATION DATE: 2023-07-31. DEVICE MANUFACTURE DATE: 2018-08-24. MEDICAL DEVICE LOT #: 8236598. MEDICAL DEVICE EXPIRATION DATE: 2023-07-31. DEVICE MANUFACTURE DATE: 2018-09-04. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BEFORE USE OF THE BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY A HAIR WAS FOUND INSIDE THE PACKAGING AND THE FLANGE WAS DAMAGED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1017122 BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS SEE H.10 30382903096054

Patients

Seq Age Sex Outcome Treatment
1 Other