FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 817863 · Received November 15, 2004

Report

Report Number
2243072-2004-00681
Event Type
Other
Date Received
November 15, 2004
Date of Event
October 17, 2004
Report Date
November 12, 2004
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT INACCURATE READINGS WITH HER PLINK METER. THE METER IS GIVING VERY ERRATIC RESULTS. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READING (378) AS REFERENCE POINT VS. BD READINGS (200, 550, 385) YIELDED DATA POINTS IN THE C ZONE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 3064293

Patients

Seq Age Sex Outcome Treatment
1 YR Other