FDA Adverse Event
Malfunction
Summary report: N
INRATIO
MDR report key: 817691
·
Received January 31, 2007
Report
- Report Number
- 2954730-2007-00055
- Event Type
- Malfunction
- Date Received
- January 31, 2007
- Date of Event
- January 29, 2007
- Report Date
- January 30, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED. DATE: 01/24/2007, 01/26/2007, INRATIO: 1.7, 1.9, LAB: 3.5, 3.8, MEAN: 2.6, 2.9, CONFIDENCE LIMITS: 1.7 - 3.8, 1.8 - 4.2. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME.
Description of Event or Problem · 1
CALLER ALLEGED INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: INRATIO: 1.7, LAB: 3.5. INRATIO: 1.9, LAB: 3.8.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | * | 060519 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |