FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 817677 · Received January 31, 2007

Report

Report Number
2954730-2007-00045
Event Type
Malfunction
Date Received
January 31, 2007
Date of Event
January 19, 2007
Report Date
January 26, 2007
Manufacturer
HEMOSENSE, INC.
Product Code
JPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INRATIO PRECISION DATA PROVIDED BY END-USER LOT: NOT AVAILABLE. FIRST TEST INR = 6.2 SECOND TEST INR = 6.1. MEAN = 6.15; SD = 0.07; %CV = 1.2%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME. DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED. DATE: 01/19/2007, INRATIO: 6.2, 6.1, LAB: 3.2, 3.2, MEAN: 4.7, 4.7, CONFIDENCE LIMITS: 2.6 - 6.9, 2.6 - 6.9. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED IMPRECISION WITH INRATIO. RESULTS AS FOLLOWS: FIRST TEST INR = 6.2 SECOND TEST INR = 6.1. CALLER ALLEGED INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE: 01/19/2007, INRATIO: 6.2, LAB: 3.2. DATE: 01/19/2007, INRATIO: 6.1, LAB: 3.2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA HEMOSENSE, INC. * NI

Patients

Seq Age Sex Outcome Treatment
1 *