FDA Adverse Event Malfunction Summary report: N

EXPRESS2 MONORAIL CORONARY STENT DELIVERY SYSTEM

MDR report key: 817651 · Received February 2, 2007

Report

Report Number
6000093-2007-00252
Event Type
Malfunction
Date Received
February 2, 2007
Date of Event
January 2, 2007
Report Date
January 4, 2007
Manufacturer
BOSTON SCIENTIFIC
Product Code
MAF
PMA / PMN Number
P020009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RETURNED PRODUCT ANALYSIS COULD NEITHER CONFIRM NOR DENY AND STENT DISLODGEMENT STATED IN THE COMPLAINT. THE ORIGINAL STENT ADHERED TO THE SNARE WIRE WAS RECEIVED AND DAMAGE WAS OBSERVED TO THE STENT. THE DELIVERY DEVICE WAS NOT RETURNED FOR ANALYSIS. VISUAL AND MICROSCOPIC EXAMINATION OF THE DAMAGED STENT DID NOT REVEAL ANY INHERENT MATERIAL DEFICIENCIES THAT WOULD HAVE CONTRIBUTED TO THE DISLODGEMENT OR STENT DAMAGE. AS THE DELIVERY DEVICE WAS NOT RETURNED FOR ANALYSIS, THE EXACT CAUSE FOR THE STENT DISLODGEMENT AND STENT DAMAGE COULD NOT BE DETERMINED. THE MANUFACTURING RECORDS FOR TOP ASSEMBLY BATCH 6813069 HAVE BEEN REVIEWED AND NO ISSUES OR DISCREPANCIES WERE FOUND. THIS RECORDS REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PERFORMANCE SPECIFICATIONS. THE ROOT CAUSE OF THE STENT DISLODGEMENT AND STENT DAMAGE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PTCA PROCEDURE, A STENT DISLODGEMENT OCCURRED. THE CALCIFIED AND TORTUOUS LESION WAS LOCATED IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THE LESION WAS PREDILATED. THE PHYSICIAN EXPERIENCED DIFFICULTY CROSSING THE LESION (RESISTANCE) AND DIFFICULTY DURING WITHDRAWAL. WHILE ATTEMPTING TO WITHDRAW THE STENT DELIVERY SYSTEM, THE STENT DISLODGED. THE STENT WAS SUCCESSFULLY REMOVED WITH A SNARE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DIFFERENT DEVICE. PATIENT STATUS IS REPORTED "OKAY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXPRESS2 MONORAIL CORONARY STENT DELIVERY SYSTEM MAF STENT, CORONARY MAF BOSTON SCIENTIFIC NA 6813069

Patients

Seq Age Sex Outcome Treatment
1 64 YR