VIPER TI SAI POLY 9X100MM
Report
- Report Number
- 1526439-2018-51151
- Event Type
- Injury
- Date Received
- December 18, 2018
- Date of Event
- January 1, 2018
- Report Date
- November 28, 2018
- Manufacturer
- MEDOS INTERNATIONAL SÃ RL CH
- Product Code
- NKB
- PMA / PMN Number
- K111571
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). A COMPLAINT INVESTIGATION WILL BE PERFORMED. THE COMPLAINT PRODUCT IS NOT AVAILABLE FOR THE INVESTIGATION. A SUPPLEMENTAL REPORT IS NOT ANTICIPATED UNLESS THE RESULTS OF THE COMPLAINT INVESTIGATION IDENTIFY A CORRECTIVE ACTION OR ADDITIONAL RELEVANT INFORMATION. SHOULD THE PRODUCT BECOME AVAILABLE, A PHYSICAL EVALUATION WILL BE CONDUCTED AND A SUPPLEMENTAL REPORT FILED WITH THE RESULTS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # ==> (B)(4). UDI: (01)UNAVAILABLE (11)UNKNOWN (10)UNKNOWN. DEVICE WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN. ALL COMPLAINT TRENDS WILL BE EVALUATED AS A PART OF THE DEPUY SPINE MONTHLY COMPLAINT REVIEW MEETING. WITHOUT THE RETURN OF THE DEVICE, WE ARE UNABLE TO CONFIRM THE REPORTED ISSUE OR IDENTIFY THE ROOT CAUSE. AS NO SYSTEMIC TRENDS WERE FOUND, THIS COMPLAINT FILE WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED. SHOULD MORE INFORMATION AND/OR THE SAMPLE BE PROVIDED AT A LATER TIME, THIS COMPLAINT WILL BE REOPENED AND DEVICE EVALUATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
(B)(4).
PRODUCT COMPLAINT # ==> (B)(4). PC: SURGICAL INTERVENTION. DEVICE IS AVAILABLE FOR EVALUATION. INVESTIGATION WILL BE CONDUCTED. FOLLOW UP WILL BE FILED WITH THE INVESTIGATION RESULTS.
PRODUCT COMPLAINT # ==> (B)(4). DEVICE IS AVAILABLE FOR EVALUATION. INVESTIGATION WILL BE CONDUCTED. FOLLOW UP WILL BE FILED WITH THE INVESTIGATION RESULTS.
PRODUCT COMPLAINT # ==> (B)(4). THE SCREW WAS BROKEN AT ITS PROXIMAL 2ND THREAD. THE BROKEN PORTION OF THE SCREW WAS NOT RETURNED. A REVIEW OF THE DEVICE HISTORY RECORD WAS CONDUCTED. NO ISSUES WERE IDENTIFIED DURING THE MANUFACTURING AND RELEASE OF THIS PRODUCT THAT COULD HAVE CONTRIBUTED TO THE PROBLEM REPORTED BY THE CUSTOMER. ALTHOUGH NO DEFINITIVE ROOT-CAUSE CAN BE DETERMINED, IT IS POSSIBLE THAT THE DEVICE EXPERIENCED UNINTENDED/EXCESSIVE FORCES SUCH AS OVER-TORQUE WHILE IN USE. AS THERE HAS BEEN NO ISSUE IDENTIFIED IN THE MANUFACTURING OR RELEASE OF THE DEVICE THAT COULD HAVE CONTRIBUTED TO THE PROBLEM REPORTED BY THE CUSTOMER, THIS COMPLAINT FILE WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
BREAKAGE OF ROD AND SAI SCREW WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1012180 | VIPER TI SAI POLY 9X100MM | ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE | NKB | MEDOS INTERNATIONAL SÃ RL CH | TBNSB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |