STERILE FX15REC W/ 4L RES
Report
- Report Number
- 1124841-2018-00330
- Event Type
- Malfunction
- Date Received
- December 17, 2018
- Date of Event
- November 29, 2018
- Report Date
- February 1, 2019
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
- Product Code
- DTZ
- UDI-DI
- 00699753450813
- PMA / PMN Number
- K151791
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS FOLLOW-UP REPORT IS SUBMITTED TO FDA IN ACCORD WITH APPLICABLE REGULATIONS ¿ AND AS INDICATED BY TERUMO CARDIOVASCULAR SYSTEMS IN THE INITIAL REPORT SUBMITTED TO THE FDA ON DECEMBER 17, 2018. (B)(4). IT WAS VERIFIED ON THE INVESTIGATION THAT THE INCORRECT OXYGENATOR WAS SHIPPED. A FAILURE IN PROCEDURE EXECUTION AS THE PICKER DID NOT PICK THE CORRECT ITEM AND THE PACKER NOR SHIPPER NOTED THE ERROR DURING THEIR VERIFICATION. IT WAS DETERMINED THAT THERE IS A DEFINED LOCATION IN THE DISTRIBUTION CENTER THAT INCLUDES PRODUCT ON PALLETS ON THE FLOOR AND THE INDICATED PRODUCT MOST LIKELY GOT INTERMINGLED WITH AN INCORRECT ITEM DURING PICKING. THE ROOT CAUSE WAS DISTRIBUTION CENTER MIXED EAST AND WEST CONFIGURED PRODUCTS WHEN RECEIVING AND PICKING ORDERS. ALL AVAILABLE INFORMATION HAS BEEN PLACED ON FILE IN QUALITY MANAGEMENT FOR APPROPRIATE TRACKING, TRENDING, AND FOLLOW-UP.
TERUMO HAS NOT RECEIVED THE DEVICE FOR EVALUATION; THEREFORE, THE INVESTIGATION HAS YET TO BE COMPLETED. TERUMO PLANS ON SUBMITTING A FOLLOW-UP REPORT WHEN THE INVESTIGATION IS COMPLETE AND WHEN MORE INFORMATION BECOMES AVAILABLE. (B)(4).
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR THAT DURING OUT OF BOX, THE CUSTOMER ORDERED "WEST ORIENTATION" OXYGENATORS, BUT THEY RECEIVED "EAST ORIENTATION" OXYGENATORS. NO PATIENT INVOLVEMENT.