Narratives

Additional Manufacturer Narrative · 0

510(K) NUMBER IS: K162717. THIS SPECIFIC EVOLUTION DEVICE IS CURRENTLY NOT REGISTERED FOR SALE IN THE US. HOWEVER, THIS DEVICE IS CONSIDERED 'SIMILAR' TO OTHER METAL BILIARY STENTS/SETS (EVOLUTION) DEVICES CURRENTLY MARKETED IN THE US. THE 510(K) NUMBER PROVIDED IS OF THE DEVICE CONSIDERED "SIMILAR" COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G. 1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(6). COOK MEDICAL INCORPORATED (CMI), 1025 ACUFF ROAD, P.O BOX 4195, BLOOMINGTON, INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: (B)(4). INVESTIGATION IS STILL PENDING, A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Additional Manufacturer Narrative · 0

510(K) NUMBER IS: K162717. THIS SPECIFIC EVOLUTION DEVICE IS CURRENTLY NOT REGISTERED FOR SALE IN THE US. HOWEVER, THIS DEVICE IS CONSIDERED 'SIMILAR' TO OTHER METAL BILIARY STENTS/SETS (EVOLUTION) DEVICES CURRENTLY MARKETED IN THE US. THE 510(K) NUMBER PROVIDED IS OF THE DEVICE CONSIDERED "SIMILAR." COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G. 1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(6); COOK MEDICAL INCORPORATED (CMI), 1025 ACUFF ROAD, P.O BOX 4195, BLOOMINGTON, INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: (B)(4). THE DEVICE INVOLVED IN THE COMPLAINT WAS EVALUATED IN THE LABORATORY ON 12TH DECEMBER 2018 AND RE-EVALUATED ON THE 16TH JANUARY 2019. NO DEFECT RELATED TO COMPLAINT FOUND IN THE LABORATORY. THE HANDLE AND INTRODUCER SYSTEM MOVEMENT IN FULL WORKING ORDER. STENT WAS ALREADY RELEASED. KINK VISIBLE ON THE DEVICE, POSSIBLY DUE TO TRANSPORTATION OF THE DEVICE BACK TO CIR AND HAS NO IMPACT, AS THE STENT WAS RELEASED. FOLLOWING THE LABORATORY EVALUATION ADDITIONAL INFORMATION WAS REQUESTED TO ELABORATE ON QUESTION IF THE STENT RELEASED. COMPLAINT DESCRIPTION IS THAT THE STENT COULD NOT BE RELEASED, HOWEVER THE STENT WAS RETURNED SEPARATELY. SECOND COMPLAINT FILE WAS RAISED TO ADDRESS THE KINK, OBSERVED DURING LABORATORY EVALUATION. DOCUMENTS REVIEW INCLUDING IFU REVIEW: PRIOR TO DISTRIBUTION ALL EVO-FC-R-20-25-12-E DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL CHECKS TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR EVO-FC-R-20-25-12-E DEVICE OF LOT NUMBER C1402106 ID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. THERE IS NO EVIDENCE TO SUGGEST THAT THIS ISSUE AFFECTS THE ENTIRE LOT #C1402106; UPON REVIEW OF COMPLAINTS THIS FAILURE MODE HAS NOT OCCURRED PREVIOUSLY WITH THIS LOT #C1402106. THE INSTRUCTIONS FOR USE (B)(4) WHICH ACCOMPANIES THIS DEVICE INSTRUCTS THE USER TO "¿VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF ANY ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE." THERE IS NO EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE (B)(4). ROOT CAUSE REVIEW: A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED AS THE CIRCUMSTANCES OF USE CANNOT BE REPLICATED IN THE LABORATORY. A POSSIBLE ROOT CAUSE COULD BE ATTRIBUTED TO A PATIENT ANATOMY. SUMMARY: ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. COMPLAINT IS CONFIRMED BASED ON THE CUSTOMERS TESTIMONY AS THE CLINICAL SETTING THAT COULD IMPACT ON THE FUNCTIONALITY OF THE DEVICE CANNOT BE REPLICATED IN THE LABORATORY. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Additional Manufacturer Narrative · 0

510(K) NUMBER IS: K162717. THIS SPECIFIC EVOLUTION DEVICE IS CURRENTLY NOT REGISTERED FOR SALE IN THE US. HOWEVER, THIS DEVICE IS CONSIDERED 'SIMILAR' TO OTHER METAL BILIARY STENTS/SETS (EVOLUTION) DEVICES CURRENTLY MARKETED IN THE US. THE 510(K) NUMBER PROVIDED IS OF THE DEVICE CONSIDERED "SIMILAR" COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G. 1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(4) COOK MEDICAL INCORPORATED (CMI) 1025 ACUFF ROAD P.O BOX 4195 BLOOMINGTON INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: 3005580113. INVESTIGATION IS STILL PENDING, A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Description of Event or Problem · 0

THIS FILE IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿DEPLOYMENT ISSUE THAT RESULTS IN THE EXPOSED STENT REMOVED FROM THE PATIENT WITH THE DELIVERY SYSTEM¿. THE STENT COULD NOT BE RELEASED. ADDITIONAL INFORMATION RECEIVED: WAS THE DIRECTIONAL BUTTON FULLY ENGAGED? YES. DID THE RED INDICATOR MOVE WHEN THE TRIGGER WAS PRESSED? YES. DID THE RED INDICATOR CONTINUE TO MOVE AFTER THE DELIVERY STOPPED? NO. WERE ANY CRACKING/POPPING SOUNDS HEARD FROM THE HANDLE? YES. WAS THE STENT PARTIALLY EXPOSED? YES. IF THE STENT WAS PARTIALLY EXPOSED, WAS IT POSSIBLE TO RECAPTURE THE STENT FULLY BEFORE REMOVAL? NO. WERE ANY ADDITIONAL PROCEDURES NEEDED? NO.

Description of Event or Problem · 0

THIS FILE IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿DEPLOYMENT ISSUE THAT RESULTS IN THE EXPOSED STENT REMOVED FROM THE PATIENT WITH THE DELIVERY SYSTEM.¿ THE STENT COULD NOT BE RELEASED. ADDITIONAL INFORMATION RECEIVED: WAS THE DIRECTIONAL BUTTON FULLY ENGAGED? YES. DID THE RED INDICATOR MOVE WHEN THE TRIGGER WAS PRESSED? YES. DID THE RED INDICATOR CONTINUE TO MOVE AFTER THE DELIVERY STOPPED? NO. WERE ANY CRACKING/POPPING SOUNDS HEARD FROM THE HANDLE? YES. WAS THE STENT PARTIALLY EXPOSED? YES. IF THE STENT WAS PARTIALLY EXPOSED, WAS IT POSSIBLE TO RECAPTURE THE STENT FULLY BEFORE REMOVAL? NO. WERE ANY ADDITIONAL PROCEDURES NEEDED? NO.

Description of Event or Problem · 0

THIS FILE IS FDA MDR REPORTABLE BASED ON THE DEVICE MALFUNCTION REPORTING PRECEDENCE FOR THIS DEVICE FAMILY FOR THE ISSUE OF ¿DEPLOYMENT ISSUE THAT RESULTS IN THE EXPOSED STENT REMOVED FROM THE PATIENT WITH THE DELIVERY SYSTEM¿. THE STENT COULD NOT BE RELEASED. ADDITIONAL INFORMATION RECEIVED: WAS THE DIRECTIONAL BUTTON FULLY ENGAGED? YES. DID THE RED INDICATOR MOVE WHEN THE TRIGGER WAS PRESSED? YES. DID THE RED INDICATOR CONTINUE TO MOVE AFTER THE DELIVERY STOPPED? NO. WERE ANY CRACKING/POPPING SOUNDS HEARD FROM THE HANDLE? YES. WAS THE STENT PARTIALLY EXPOSED? YES. IF THE STENT WAS PARTIALLY EXPOSED, WAS IT POSSIBLE TO RECAPTURE THE STENT FULLY BEFORE REMOVAL? NO. WERE ANY ADDITIONAL PROCEDURES NEEDED? NO.