FDA Adverse Event Malfunction Summary report: N

BD POSIFLUSH¿ NS FILLED SYRINGE

MDR report key: 8166373 · Received December 14, 2018

Report

Report Number
9616657-2018-00068
Event Type
Malfunction
Date Received
December 14, 2018
Date of Event
November 25, 2018
Report Date
December 19, 2018
Manufacturer
BECTON, DICKINSON AND CO.
Product Code
NGT
UDI-DI
30382903065463
PMA / PMN Number
K141311
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: ONE SAMPLE BELONGING TO LOT NUMBER 8192991 WAS RECEIVED FOR EVALUATION BY OUR QUALITY ENGINEER. THROUGH VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED SAMPLE, IT WAS DETERMINED THAT THE PLUNGER MOVEMENT WAS DEFECTIVE. DHR REVIEW SHOWS THAT DURING THE PRODUCTION PROCESS OF LOT NUMBER 8192991, THERE WAS ONE DOCUMENTED NON-CONFORMANCE THAT COULD HAVE CONTRIBUTED TO THIS DEFECT. DURING PRODUCTION, A DAMAGED SILICONE CANISTER WAS FOUND WITHIN THE SILICONE STATION, WHICH RESULTED IN AN INSUFFICIENT SUPPLY OF SILICONE TO THE SYRINGE BARRELS. PRODUCT ASSOCIATED WITH THIS NON-CONFORMANCE WAS HELD FOR INSPECTION AND ALL AFFECTED MATERIAL SHOULD HAVE BEEN DISCARDED. IT IS POSSIBLE THAT A MISDETECTION IN DEFECTIVE PRODUCT CAUSED THIS INCIDENT, AS THE SAMPLES WERE NOT PROPERLY DISCARDED. THE DAMAGED SILICONE CANISTER HAS SINCE BEEN REPLACED. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PLUNGER IN THE BD POSIFLUSH¿ NS FILLED SYRINGE WOULDN'T MOVE.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PLUNGER IN THE BD POSIFLUSH¿ NS FILLED SYRINGE WOULDN'T MOVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1003103 BD POSIFLUSH¿ NS FILLED SYRINGE INTRAVASCULAR CATHETER NGT BECTON, DICKINSON AND CO. 8192991 30382903065463

Patients

Seq Age Sex Outcome Treatment
1 Other